One-year outcomes of the BeGraft stent graft used as chimney graft in conjunction with the Endurant device for the treatment of complex abdominal diseases
Aim To evaluate the performance of two different flexible devices, the balloon-expandable BeGraft stent as on-label chEVAR use with the Endurant (Medtronic, Santa Rosa, USA) stent graft compared to the self-expandable off-label employment of Viabahn (Gore Flagstaff, Arizona, USA). Methods All patients treated for pararenal aortic pathologies between 2008 and 2017 using Endurant II (Medtronic, Santa Rosa, CA, USA) abdominal endograft in combination with Viabahn (Gore, Flagstaff, AZ, USA) or BeGraft (Bentley, Hechingen, Germany) at our institution were retrospectively analyzed. Patients were divided into two groups according to the employed chimney graft, namely Viabahn (group A) and BeGraft (group B). The primary endpoint was the incidence of type IA endoleak at one month and last follow-up. Secondary endpoints included reintervention rates, primary patency of the chimney graft, and need for additional placement of bare metal stents also known as relining. Results A total of 27 patients (mean age: 76 years) were treated with overall 47 chimney grafts. In particular, 11 patients received 21 Viabahn and 16 patients underwent placement of 26 BeGraft stents. Mean follow-up was 14.3 ± 9.0 months for group A and 13.0 ± 13.7 for group B, p = .451. The mean chimney graft length was 85.1 ± 38.9 mm for group A and 34.9 ± 9.8mm for group B ( p < .001). The oversizing rate at the proximal sealing zone was significantly higher in group B (Group A: 23.1 ± 16.5% vs. Group B: 33.6 ± 7.6%, p = .007). The primary patency rates were similar between groups ( p = .250). The subgroup analysis showed a significantly higher early (9–42.9% vs. 2–7.7%, p = .005) and at last CT follow-up (5–23.8% vs. 0–0.0%, p = .011) type IA endoleak, higher primary relining (14–66.7% vs. 5–19.2%, p = .001) and reintervention rates (5–23.8% vs. 0–0.0%; p = .011) for group A chimney stent grafts. Conclusions Both chEVAR combinations displayed similar patency rate, showing the BeGraft-Endurant composition less need for relining and fewer one-year type IA endoleak and reintervention rates in the subgroup analysis.