scholarly journals The Influence of Plasticizers on Determination of Cationic Surfactants in Pharmaceutical Disinfectants by Direct Potentiometric Surfactant Sensor

2020 ◽  
Vol 2 (1) ◽  
pp. 44
Author(s):  
Maja Karnaš ◽  
Marija Jozanović ◽  
Bojan Đurin ◽  
Nikola Sakač

Ion selective liquid membrane type membranes are usually made of a PVC, ionophore and a plasticizer. Plasticizers soften the PVC but due to their lipophilicity they influence the ionophore, the ion exchange across the membrane, membrane resistance and consequently the analytical signal. The aim of the research was to investigate the influence of four different plasticizers, in formulation with the same ionophore, on the analytical properties of the sensor membrane towards two often used cationic surfactants, then select the best membrane formulation and test it on real samples of six pharmaceutical disinfectants containing cationic surfactants.

2003 ◽  
Vol 1 (3) ◽  
pp. 242-259 ◽  
Author(s):  
Ewa Gorodkiewicz ◽  
Paweł Falkowski ◽  
Anna Sankiewicz ◽  
Zbigniew Figaszewski

AbstractThe analytical properties of an ion-selective electrode sensitive to labetalol with a liquid membrane, based on ion-pair complexes with sodium tetraphenylborate (TPB-Na+) are described. The studied electrode can be used for the determination of labetalol hydrochloride as a protonated form of labetalol in pharmaceuticals. The calibration curve, e.g. EMF=f(pCLabHCl) is linear in the range from 10−5 to 10−2 mol L−1 with a correlation coefficient of 0.9992 and slope of 61.13 mV/decade, which is close to the Nernstian slope. The detection limit of the examined electrode is 7.20×10−6 mol L−1. The influence of pH of the tested solutions on the formulation of the electrode is not as considerable since the electrode works correctly in the pH range 3.0–8.0. The main attributes of the developed electrode are: stability, good reproducibility of EMF and short response time, close to 30 seconds depending on labetalol concentration in the solution. The electrode shows good selectivity for many inorganic ions. The selectivity for drug cations is weaker due to the structural similarity of the interfering cations to labetalol. The results of labetalol determination using direct potentiometry in drugs such as Pressocard (Polpharma) and Trandate (GlaxoWellcome) were compatible with the quantity of labetalol declared by the manufacturer, and with parallel UV spectrophotometric and HPLC determinations.


Sensors ◽  
2021 ◽  
Vol 21 (10) ◽  
pp. 3535
Author(s):  
Nikola Sakač ◽  
Dubravka Madunić-Čačić ◽  
Maja Karnaš ◽  
Bojan Đurin ◽  
Ivan Kovač ◽  
...  

Surfactant liquid-membrane type sensors are usually made of a PVC, ionophore and a plasticizer. Plasticizers soften the PVC. Due to their lipophilicity, they influence the ion exchange across the membrane, ionophore solubility, membrane resistance and, consequently, the analytical signal. We used the DMI-TPB as an ionophore, six different plasticizers [2-nitrophenyl-octyl-ether (P1), bis(2-ethylhexyl) phthalate (P2), bis(2-ethylhexyl) sebacate (P3), 2-nitrophenyl phenyl ether (P4), dibutyl phthalate (P5) and dibutyl sebacate (P6)] and a PVC to produce ionic surfactant sensors. Sensor formulation with P1 showed the best potentiometric response to four usually used cationic surfactant, with the lowest LOD, 7 × 10−7 M; and potentiometric titration curves with well-defined and sharp inflexion points. The sensor with P6 showed the lowest analytical performances. Surfactant sensor with P1 was selected for quantification of cationic surfactant in model solutions and commercial samples of disinfectants and antiseptics. It showed high accuracy and precision in all determinations, with recovery from 98.2 to 99.6, and good agreement with the results obtained with surfactant sensor used as a referent one, and a standard two-phase titration method. RDS values were lower than 0.5% for all determinations.


2018 ◽  
Vol 84 (11) ◽  
pp. 9-14
Author(s):  
E. S. Koshel ◽  
V. B. Baranovskaya ◽  
M. S. Doronina

The analytical capabilities of arc atomic emission determination of As, Bi, Sb, Cu, Te in rare earth metals (REM) and their oxides after preparatory group concentration using S,N-containing heterochain polymer sorbent are studied on a high-resolution spectrometer “Grand- Extra” (“WMC-Optoelectron-ics” company, Russia). Sorption kinetics and dependence of the degree of the impurity extraction on the solution acidity are analyzed to specify conditions of sorption concentration. To optimize the procedure of arc atomic emission determination of As, Bi, Sb, Cu, and Te various schemes of their sorption preconcentration and subsequent processing of the resulted concentrate with the addition of a collector at different stages of the sorption process have been considered. Graphite powder is used as a collector in analysis of rare earth oxides due to universality and relative simplicity of the emission spectrum. Conditions of analysis and parameters of the spectrometer that affect the analytical signal (mass and composition of the sample, shape and size of the electrodes, current intensity and generator operation mode, interelectrode spacing, wavelengths of the analytical lines) are chosen. The evaporation curves of the determinable impurities were studied and the exposure time of As, Bi, Sb, Cu, and Te in the resulted sorption concentrate was determined. Correctness of the obtained results was evaluated using standard samples of the composition and in comparisons between methods. The results of the study are used to develop a method of arc chemical-atomic emission analysis of yttrium, gadolinium, neodymium, europium, scandium and their oxides in a concentration range of n x (10-2 - 10-5) wt.%.


2020 ◽  
Vol 16 (7) ◽  
pp. 872-879
Author(s):  
Samin Hamidi

Background: Abuse of drugs is associated with several medical, forensic, toxicology and social challenges. “Drugs of abuse” testing is therefore an important issue. Objective: We propose a simple CE-based method for the quantification of amphetamine, codeine and morphine after direct injection of Exhaled Breath Condensate (EBC) by the aid of simple stacking mode and an off-line pre-concentration method. Methods: Using graphene oxide adsorbents, amphetamine, codeine and morphine were extracted from EBC in order to eliminate the proteins and other interferences. In addition to off-line method, an online stacking mode was applied to improve the analytical signal obtained from the instrument. Results: The validation parameters were experimented on the developed method based on the FDA guideline over concentration ranges of 12.5-100, 30-500 and 10-1250 ng/mL associated with amphetamine, codeine and morphine, respectively. Small volumes (around 100 μL) of EBC were collected using a lab-made setup and successfully analyzed using the proposed method where precisions and accuracies (within day and between days) were in accordance with the guideline (recommended less than 15 % for biological samples). The recovery tests were used to evaluate the matrix effect and data (94 to 105 %) showed that the proposed method can be applied in different EBC matrix samplings of subjects. Conclusion: The proposed method is superior for simultaneous determination of amphetamine, codeine and morphine over chromatographic analyses because it is fast and consumes fewer chemicals, with no derivatization step.


1984 ◽  
Vol 49 (10) ◽  
pp. 2349-2354 ◽  
Author(s):  
František Vláčil ◽  
Karel Koňák

The selectivity coefficients of the nitrate and chloride ions and of anionic chloro complexes of Au(III), Rh(III), Pd(II), and Pt(IV) for ion exchange on Spheron DEAE in the chloride form are determined by the dynamic method. the complex anion species formed are identified and the ion exchange nature of the sorption of precious metals on this sorbent is confirmed based on the elution order of the precious metals as determined previously by the column chromatography on Spheron DEAE using hydrochloric acid as the mobile phase. The effect of the presence of perchlorate in the mobile phase during the liquid chromatography of precious metals and during the chromatographic determination of nitrate traces is explained.


1992 ◽  
Vol 57 (9) ◽  
pp. 1905-1914
Author(s):  
Miroslav Bleha ◽  
Věra Šumberová

The equilibrium sorption of uni-univalent electrolytes (NaCl, KCl) in heterogeneous cation exchange membranes with various contents of the ion exchange component and in ion exchange membranes Ralex was investigated. Using experimental data which express the concentration dependence of equilibrium sorption, validity of the Donnan relation for the systems under investigation was tested and values of the Glueckauf inhomogeneity factor for Ralex membranes were determined. Determination of the equilibrium sorption allows the effect of the total content of internal water and of the ion-exchange capacity on the distribution coefficients of the electrolyte to be determined.