scholarly journals Effect of the Acid-Etching on Grit-Blasted Dental Implants to Improve Osseointegration: Histomorphometric Analysis of the Bone-Implant Contact in the Rabbit Tibia Model

Coatings ◽  
2021 ◽  
Vol 11 (11) ◽  
pp. 1426
Author(s):  
Blanca Ríos-Carrasco ◽  
Bernardo Ferreira Lemos ◽  
Mariano Herrero-Climent ◽  
F. Javier Gil Mur ◽  
Jose Vicente Ríos-Santos

Previous studies have shown that the most reliable way to evaluate the success of an implant is by bone-to-implant contact (BIC). Recent techniques allow modifications to the implant surface that improve mechanical and biological characteristics, and also upgrade osseointegration. Objective: The aim was to evaluate the osseointegration in rabbit tibia of two different titanium dental implant surfaces: shot-blasted with Al2O3 (SB) and the same treatment with an acid-etching by immersion for 15 s in HCl/H2SO4 (SB + AE). Material and methods: Roughness parameters (Ra, Rt, and Rz) were determined by white light interferometer microscopy. Surface wettability was evaluated with a contact angle video-based system using water, di-iodomethane, and formamide. Surface free energy was determined by means of Owens and Wendt equations. Scanning electron microscopy equipped with X-ray microanalysis was used to study the morphology and determine the chemical composition of the surfaces. Twenty-four grade 4 titanium dental implants (Essential Klockner®) were implanted in the rabbit’s tibia, 12 for each surface treatment, using six rabbits. Six weeks later the rabbits were sacrificed and the implants were sent for histologic analysis. Resonance frequency analysis (RFA) was recorded both at the time of surgery and the end of the research with each device (Osstell Mentor and Osstell ISQ). Results: The roughness measurements between the two treatments did not show statistically significant differences. However, the effect of the acid etching made the surface slightly more hydrophilic (decreasing contact angle from 74.7 for SB to 64.3 for SB + AE) and it presented a higher surface energy. The bone-to-implant contact ratio (BIC %) showed a similar tendency, with 55.18 ± 15.67 and 59.9 ± 13.15 for SB and SB + AE implants, respectively. After 6 weeks of healing, the SB + AE showed an implant stability quotient (ISQ) value of 76 ± 4.47 and the shot-blasted one an ISQ value of 75.83 ± 8.44 (no statistically significant difference). Implants with different surface properties had distinctive forms of behavior regarding osseointegration. Furthermore, the Osstell system was an invasive and reliable method to measure implant stability. Conclusion: Both surfaces of implants studied showed high osseointegration. The SB and SB + AE implants used in our study had similar behavior both in terms of BIC values and RFA. The RFA systems in Osstell Mentor and Osstell ISQ confirmed nearly perfect reproducibility and repeatability.

2018 ◽  
Vol 2018 ◽  
pp. 1-8 ◽  
Author(s):  
Sergio Alexandre Gehrke ◽  
Raphaél Bettach ◽  
Jaime Sardá Aramburú Júnior ◽  
Juan Carlos Prados-Frutos ◽  
Massimo Del Fabbro ◽  
...  

Objectives. The present study aims to compare the drilling protocol effect on osseointegration event in three commercially available titanium dental implants with different drill protocol using a rabbit tibia model. Materials and Methods. Three different drilling sequences were compared as follows: drilling sequence using a single unique drill of 4.2 mm conical implant (Group 1), drilling sequence using 3 consecutive cylindrical drills for a 4.1 mm cylindrical implant (Group 2), and drilling sequence using 3 consecutive conical drills for a 4.3 mm conical implant (Group 3). For each group, 18 drilling procedures and implant placements were performed, totalizing 54 commercially available titanium dental implants. The samples were removed 6 weeks after implantation. Resonance frequency analyses (RFA) were performed immediately after the implantation, and at 6 weeks removal torque test (RTt) and histological analysis were performed. Results. The RFA measured showed statistical difference between the groups in time 1 and no significant statistical differences in time 2 (p>0.05). In the RTt no significant difference was found between the 3 groups tested. Histomorphometric analysis showed no significant difference between groups in the bone-to-implant contact% (p>0.05). Conclusion. In the present preclinical study, osteotomy using a single bur did not show differences regarding the proposed and evaluated tests parameters for assessing the peri-implant behavior.


2021 ◽  
Vol 25 (1) ◽  
Author(s):  
Hao-Zhen Lyu ◽  
Jae Hyup Lee

Abstract Background Histology is considered as a gold standard for analyzing bone architecture. However, histomorphometry is a destructive method and only offers the bone information of a limited location. Micro-computed tomography (μCT) is a non-destructive technology and provides a slice at any site. The aim of this study was to compare the correlation of the Bone-to-Implant Contact ratio (BIC) between 2D micro-CT (μCT) and histomorphometry and to investigate a method for assessing the osseointegration of the implant by 2D μCT. Methods A total of 18 implants were divided into three groups (6 implants per group), and inserted into the rabbit tibia defects as follow: implant only (Implant group), implant with β-TCP/hydrogel (TCP group), implant with rhBMP-2 loaded β-TCP/hydrogel composite (BMP-2 group). After 4 weeks of implantation, the specimens were collected to take the micro-CT scan with an aluminum filter and performed H&E staining on the undecalcified sections. The 2D μCT slices were chosen at an angle of 0°, 45°, 90° and 135° with the representative histological section to measure BIC. And the correlations between BICs of 2D μCT and BICs of histology were evaluated. Results In each group, BICs at the same sites measured by histomorphometry and corresponding 2D μCT presented the same trend and shown no significant difference between the two methods (P > 0.05). BICs of histological sections and BICs of corresponding 2D μCT slices presented a strong correlation in the implant group (γ = 0.74, P = 0.09), a moderate correlation in the TCP group (γ = 0.46, P = 0.35), a weak correlation in the BMP-2 group (γ = 0.30, P = 0.56). In the implant group, the relationship between BIC-Mean-μCTs and BICs-Histology has presented a significant linear correlation (γ = 0.84, P = 0.04). Conclusions Integrating bone information of several 2D μCT slices in different sites to measure BIC is a feasible method for assessing the implant osseointegration.


Materials ◽  
2019 ◽  
Vol 12 (13) ◽  
pp. 2078 ◽  
Author(s):  
Jun-Beom Lee ◽  
Ye-Hyeon Jo ◽  
Jung-Yoo Choi ◽  
Yang-Jo Seol ◽  
Yong-Moo Lee ◽  
...  

Ultraviolet (UV) photofunctionalization has been suggested as an effective method to enhance the osseointegration of titanium surface. In this study, machined surface treated with UV light (M + UV) was compared to sandblasted, large-grit, acid-etched (SLA) surface through in vitro and in vivo studies. Groups of titanium specimens were defined as machined (M), SLA, and M + UV for the disc type, and M + UV and SLA for the implant. The discs and implants were assessed using scanning electron microscopy, confocal laser scanning microscopy, electron spectroscopy for chemical analysis, and the contact angle. Additionally, we evaluated the cell attachment, proliferation assay, and real-time polymerase chain reaction for the MC3T3-E1 cells. In a rabbit tibia model, the implants were examined to evaluate the bone-to-implant contact ratio and the bone area. In the M + UV group, we observed the lower amount of carbon, a 0°-degree contact angle, and enhanced osteogenic cell activities (p < 0.05). The histomorphometric analysis showed that a higher bone-to-implant contact ratio was found in the M + UV implant at 10 days (p < 0.05). In conclusion, the UV photofunctionalization of a Ti dental implant with M surface attained earlier osseointegration than SLA.


2016 ◽  
Vol 42 (5) ◽  
pp. 445-450 ◽  
Author(s):  
Dennis Flanagan

After dental implants are manufactured there can be a loss of biological activity that may be reactivated by exposure to ultraviolet (UV) radiation, that is, photofunctionalization. The titanium surface is energy conditioned by UV radiation. This imparts a slight positive surface energy and hydrophilicity to the titanium dental implant surface. This conditioning renews biological activity lost after a shelf life of as little as 2 weeks. The UV radiation has chemical and biological effects on the osseous-implant interface. Photofunctionization for as little as 15 minutes accelerates healing and increases bone to implant contact. The most effective time exposure and UV wave length are in need of identification to produce a surface most conducive for osseointegration.


Author(s):  
Sigmar Kopp ◽  
Mareike Warkentin ◽  
Ferenc Öri ◽  
Peter Ottl ◽  
Günther Kundt ◽  
...  

AbstractThis study was designed to determine and statistically analyze bone-to-implant contact (BIC) values for human specimens segmented in at least two different locations.Samples of human bone with fractured osseointegrated implants were obtained from six patients. Sections were prepared, dehydrated, and resin infiltrated. Undecalcified bone sections were produced using the thin-section technique according to Donath, ultimately obtaining a section thickness of approximately 20 μm. Fifteen specimens were available for histomorphometry. The bone sections were digitized and analyzed. The bone-to-metal contact (BMC) parameter was determined histomorphometrically. The BMC was returned in terms of the visibly bone-covered implant surfaces as a percentage of the total implant surface shown.The values obtained for the six implants were arranged as six maximum-distance pairs and tested for significance using the t-test for dependent samples. The mean difference in BIC was 11.69±9.79%. The two-sided test showed a significant difference (p=0.033).The accidental or deliberate choice of section plane for a bone-implant block has an influence on the BIC value. To make BIC values more comparable, a standardization of section planes is desirable.


Author(s):  
Karthikeyan Subramani

This manuscript reviews about titanium surface modification techniques for its application in orthopaedic and dental implants. There are a few limitations in the long term prognosis of orthopaedic and dental implants. Poor osseointegration with bone, periimplant infection leading to implant failure and short term longevity demanding revision surgery, are to mention a few. Micro- and nanoscale modification of titanium surface using physicochemical, morphological and biochemical approaches have resulted in higher bone to implant contact ratio and improved osseointegration. With recent advances in micro, nano-fabrication techniques and multidisciplinary research studies focusing on bridging biomaterials for medical applications, TiO2 nanotubes have been extensively studied for implant applications. The need for titanium implant surface that can closely mimic the nanoscale architecture of human bone has become a priority. For such purpose, TiO2 nanotubes of different dimensions and architectural fashions at the nanoscale level are being evaluated. This manuscript discusses in brief about the in-vitro and in-vivo studies on titanium surface modification techniques. This manuscript also addresses the recent studies done on such nanotubular surfaces for the effective delivery of osteoinductive growth factors and anti bacterial/ anti inflammatory drugs to promote osseointegration and prevent peri-implant infection.


2012 ◽  
Vol 2012 ◽  
pp. 1-7 ◽  
Author(s):  
Carlos Nelson Elias ◽  
Patricia Abdo Gravina ◽  
Costa e Silva Filho ◽  
Pedro Augusto de Paula Nascente

Statement of Problem. The chemical or topographic modification of the dental implant surface can affect bone healing, promote accelerated osteogenesis, and increase bone-implant contact and bonding strength.Objective. In this work, the effects of dental implant surface treatment and fibronectin adsorption on the adhesion of osteoblasts were analyzed.Materials and Methods. Two titanium dental implants (Porous-acid etching and PorousNano-acid etching followed by fluoride ion modification) were characterized by high-resolution scanning electron microscopy, atomic force microscopy, and X-ray diffraction before and after the incorporation of human plasma fibronectin (FN). The objective was to investigate the biofunctionalization of these surfaces and examine their effects on the interaction with osteoblastic cells.Results. The evaluation techniques used showed that the Porous and PorousNano implants have similar microstructural characteristics. Spectrophotometry demonstrated similar levels of fibronectin adsorption on both surfaces (80%). The association indexes of osteoblastic cells in FN-treated samples were significantly higher than those in samples without FN. The radioactivity values associated with the same samples, expressed as counts per minute (cpm), suggested that FN incorporation is an important determinant of thein vitrocytocompatibility of the surfaces.Conclusion. The preparation of bioactive titanium surfaces via fluoride and FN retention proved to be a useful treatment to optimize and to accelerate the osseointegration process for dental implants.


2012 ◽  
Vol 91 (12) ◽  
pp. 1172-1177 ◽  
Author(s):  
R. Jimbo ◽  
P.G. Coelho ◽  
M. Bryington ◽  
M. Baldassarri ◽  
N. Tovar ◽  
...  

Nanostructure modification of dental implants has long been sought as a means to improve osseointegration through enhanced biomimicry of host structures. Several methods have been proposed and demonstrated for creating nanotopographic features; here we describe a nanoscale hydroxyapatite (HA)-coated implant surface and hypothesize that it will hasten osseointegration and improve its quality relative to that of non-coated implants. Twenty threaded titanium alloy implants, half prepared with a stable HA nanoparticle surface and half grit-blasted, acid-etched, and heat-treated (HT), were inserted into rabbit femurs. Pre-operatively, the implants were morphologically and topographically characterized. After 3 weeks of healing, the samples were retrieved for histomorphometry. The nanomechanical properties of the surrounding bone were evaluated by nanoindentation. While both implants revealed similar bone-to-implant contact, the nanoindentation demonstrated that the tissue quality was significantly enhanced around the HA-coated implants, validating the postulated hypothesis.


2018 ◽  
Author(s):  
Preeti Satheesh Kumar ◽  
Vyoma Venkatesh Grandhi ◽  
Vrinda Gupta

BACKGROUND . A variety of claims are made regarding the effects of surface topography on implant osseointegration. The development of implant surfaces topography has been empirical, requiring numerous in vitro and in vivo tests. Most of these tests were not standardized, using different surfaces, cell populations or animal models. The exact role of surface chemistry and topography on the early events of the osseointegration of dental implants remain poorly understood. OBJECTIVE This review considers the major claims made concerning the effects of titanium implant surface topography on osseointegration. The osseointegration rate of titanium dental implants is related to their composition and surface roughness. The different methods used for increasing surface roughness or applying osteoconductive coatings to titanium dental implants are reviewed. Important findings of consensus are highlighted, and existing controversies are revealed. METHODS This review considers many of the research publications listed in MEDLINE and presented in biomedical research publications and textbooks. Surface treatments, such as titanium plasma-spraying, grit-blasting acid-etching,alkaline etching, anodization,polymer demixing ,sol gel conversion and their corresponding surface morphologies and properties are described. RESULTS Many in vitro evaluations are not predictive of or correlated with in vivo outcomes. In some culture models, increased surface topography positively affects pro-osteogenic cellular activities. Many studies reveal increase in bone-to-implant contact,with increased surface topography modifications on implant surfaces. CONCLUSIONS Increased implant surface topography improves the bone-to-implant contact and the mechanical properties of the enhanced interface.