Comparative Evaluation of Hematologic Analyzers Advia® 120, Advia® 2120 and Pentra® 120DX for Platelet Counts below 40 X 109/L.

Blood ◽  
2005 ◽  
Vol 106 (11) ◽  
pp. 1256-1256
Author(s):  
Josep M. Jou ◽  
Fulgencio Navalon ◽  
Ester Jiménez ◽  
Maribel Diaz-Ricart ◽  
Rosa Brugues ◽  
...  

Abstract More aggressive therapies used for treatment of oncohematological malignancies or control of immune responses are resulting in an increased frequency of platelet counts below the 50 x 109/L limit. The recommended reference method for platelet counts was tedious and showed low reproducibility until now. In the last 2 years, flow cytometry based techniques combined with specific monoclonal antibodies (MoAbs) have been accepted as reference method. We have evaluated the accuracy for low platelet counts of several hematologic analyzers currently used in our laboratories. The new reference method approved by ISLH, ICSH y NCCLS is based on double labeling of platelets using MoAbs directed to CD41 and CD61 followed by flow cytometry analysis. Absolute platelet counts are calculated using a ratio with red blod cell (RBC) counts provided by the hematological analyzers. In our studies, 50 blood samples with platelet counts ranging from 1.5 to 39.4 x109/L were processed in duplicate through 1 Advia 2120 (Bayer Diagnostics), 2 Advia 120 (Bayer Diagnostics) and 2 Pentra 120 DX (Horiba-ABX Diagnostics). Advia analyzers use laser-based technology while Pentra analyzers use impedance one for cell counting. All samples were also processed through the reference flow cytometric method, being platelets identified by their double labeling for CD41 and CD61. The minimal number of platelets acquired in the platelet region was established at 1000. Absolute platelet counts were calculated using RBC counts provided by the respective analyzers. All blood samples were processed within 6 hours from phlebotomy. Statistical methods applied included: coefficient of variation (%CV), coefficient of correlation (r ), linear regression, Passing-Bablock (P-B) regression and Bland-Altman test. Precision of each analyzer was obtained by processing in 10 times different blood samples with counts from 4 to 39 x 109/L. Global results were evaluated, though special attention was paid to subgroups of results below or above 20 x 109/L. Correlation between reference values and counts provided by the Advia 2120 was 0.945 with a linear regression of 0.987x+2.9. P-B correlation was good and the average difference was 2.7 x 109/L. In the subgroup of samples with counts below 20 x 109/L correlation was 0.874 with 1.00x+2.7. P-B was correct and the average difference was 2.8 x 109/L. Results with Advia 120 were always similar to those calculated with the Advia 2120, though the average difference was slightly lower with a value of 1.7 x 109/L. Precision (CV) was 16% for platelet count levels at 4 x 109/L, 12% for those at 13 x 109/L and 4% for those at 39 x 109/L. Correlation with Pentra 120 Dx was 0.937 with a linear regression of 0.894x+2.7, the P-B was acceptable with an average difference of 1.2 x 109/L. Correlation index was 0.824 with a linear regression of 0.88x+2.8 for platelet counts below 20 x 109/L, average difference was of 1.4 x 109/L and a correct P-B. Precision (CV) ranged from 26% at 4 x 109/L and 8% at 20 x 109/L platelet counts. Our data demonstrate that hematological analyzers evaluated provided very reliable results at low platelet counts. Advia and Pentra analyzers investigated tend to over calculate the number of platelets (2.5 and 1.4 x109/L respectively). Correlation scattering was slightly superior with the Pentra analyzer. Overall reproducibility for low platelet counts was excellent for both laser and impedance technologies tested.

PLoS ONE ◽  
2021 ◽  
Vol 16 (8) ◽  
pp. e0256423
Author(s):  
William M. Dickerson ◽  
Rebecca Yu ◽  
Helena U. Westergren ◽  
Jonathan Paraskos ◽  
Philipp Schatz ◽  
...  

Background Point-of-care (PoC) testing of platelet count (PLT) provides real-time data for rapid decision making. The goal of this study is to evaluate the accuracy and precision of platelet counting using a new microvolume (8 μL), absolute counting, 1.5 kg cytometry-based blood analyzer, the rHEALTH ONE (rHEALTH) in comparison with the International Society of Laboratory Hematology (ISLH) platelet method, which uses a cytometer and an impedance analyzer. Methods Inclusion eligibility were healthy adults (M/F) ages 18–80 for donation of fingerprick and venous blood samples. Samples were from a random N = 31 volunteers from a single U.S. site. Samples were serially diluted to test thrombocytopenic ranges. Interfering substances and conditions were tested, including RBC fragments, platelet fragments, cholesterol, triglycerides, lipids, anti-platelet antibodies, and temperature. Results The concordance between the rHEALTH and ISLH methods had a slope = 1.030 and R2 = 0.9684. The rHEALTH method showed a correlation between capillary and venous blood samples (slope = 0.9514 and R2 = 0.9684). Certain interferents changed platelet recovery: RBC fragments and anti-platelet antibodies with the ISLH method; platelet fragments and anti-platelet antibodies on the rHEALTH; and RBC fragments, platelets fragments, triglycerides and LDL on the clinical impedance analyzer. The rHEALTH’s precision ranged from 3.1–8.0%, and the ISLH from 1.0–10.5%. Conclusions The rHEALTH method provides similar results with the reference method and good correlation between adult capillary and venous blood samples. This demonstrates the ability of the rHEALTH to provide point-of-care assessment of normal and thrombocytopenic platelet counts from fingerprick blood with high precision and limited interferences.


1989 ◽  
Vol 35 (8) ◽  
pp. 1734-1739 ◽  
Author(s):  
P S Bachorik ◽  
R H Bradford ◽  
T Cole ◽  
I Frantz ◽  
A M Gotto ◽  
...  

Abstract We compared plasma cholesterol measurements made with the Boehringer Mannheim Reflotron reflectance photometric analyzer in 1298 capillary blood samples with measurements made in venous blood samples collected at the same time and analyzed in four standardized Lipid Research Clinics laboratories. The Reflotron measurements averaged 0.8% to 7.8% lower than the laboratory values. Correlations (r) between the two sets of measurements ranged from 0.92 to 0.96. In some samples, however, the Reflotron values differed from the laboratory values by greater than or equal to 12%; the cholesterol concentrations in these samples tended to be higher than in those for which better agreement was observed. The smaller negative biases were observed when test strips were used that were calibrated with reference to the Centers for Disease Control Reference Method for cholesterol. The agreement between sequential Reflotron values averaged less than or equal to 4.3%. There was an average difference of less than or equal to 1.0% between Reflotron measurements made in each of two sequential capillary blood samples taken from a single finger puncture.


2003 ◽  
Vol 121 (1) ◽  
pp. 19-23 ◽  
Author(s):  
Raimundo Antônio Gomes Oliveira ◽  
Maria Mariko Takadachi ◽  
Kimiyo Nonoyama ◽  
Orlando César de Oliveira Barretto

CONTEXT: Reliable platelet counting is crucial for indicating prophylactic platelet transfusion in thrombocytopenic patients. OBJECTIVE: To evaluate the precision and accuracy of platelet counting for thrombocytopenic patients, using four different automated counters in comparison with the Brecher & Cronkite reference method recommended by the International Committee for Standardization in Hematology (ICSH). TYPE OF STUDY: Automated platelet counting assessment in thrombocytopenic patients. SETTING: Hematology Laboratory, Hospital do Servidor Público Estadual de São Paulo, and the Hematology Division of Instituto Adolfo Lutz, São Paulo, SP, Brazil. MAIN MEASUREMENTS: Brecher & Cronkite reference method and four different automated platelet counters. PARTICIPANTS: 43 thrombocytopenic patients with platelet counts of less than 30,000/µl RESULTS: The ADVIA-120 (Bayer), Coulter STKS, H1 System (Technicom-Bayer) and Coulter T-890 automatic instruments presented great precision and accuracy in relation to laboratory thrombocytopenic samples obtained by diluting blood from normal donors. However, when thrombocytopenic patients were investigated, all the counters except ADVIA (which is based on volume and refraction index) showed low accuracy when compared to the Brecher & Cronkite reference method (ICSH). The ADVIA counter showed high correlation (r = 0.947). However, all counters showed flags in thrombocytopenic samples. CONCLUSION: The Brecher & Cronkite reference method should always be indicated in thrombocytopenic patients for platelet counts below 30,000 plt /µl obtained in one dimensional counters.


Blood ◽  
2010 ◽  
Vol 116 (21) ◽  
pp. 4727-4727
Author(s):  
Guat Bee Tan ◽  
Christina Sum ◽  
Ponnudurai Kuperan

Abstract Abstract 4727 The examination of blood films by microscopy remains one of the major labour intensive procedures in the laboratory and the challenge is to reduce the number of blood films examined without missing important diagnostic information. Automated blood cell counters offer a leucocyte count, red cell and platelet count and five-part (some 6-part) leucocyte differential. Haematology instrument differentials provide only limited information on cell morphology using abnormal cell flags and are often unable to reliably classify abnormal and immature cells. The examination of blood films is not only time consuming, it also requires highly trained staff. The impact of a wrong diagnosis necessitates that experienced staff are present in the laboratory 24 hours a day. Furthermore, manual cell classification is subjective, with significant inter and intra observer variation (Koepke et al. 1985) and is also subject to significant statistical variance (Rumke 1985). There have recently been several reports of using monoclonal antibody cocktails for an extended leucocyte differential by flow cytometry (Faucher et al. 2007, Roussel et al. 2010). The aim of this study was to compare a flow cytometric method for the white blood cell differential with the automated count from the Beckman Coulter LH750 haematology analyser and the reference manual microscopic 2 × 200 cell count (CLSI H20-A2). Cell morphology was also assessed microscopically for the presence of cells such as reactive or abnormal lymphocytes or blasts. The flow cytometric method, described by Faucher et al. 2007, uses 6 antibodies (CD45, CD36, CD2, CD294, CD19 and CD16) premixed in a single tube. The protocol allows detection of all white blood cells, mature neutrophils, total lymphocytes, total monocytes, eosinophils, basophils, immature granulocytes, B lymphocytes, non-cytotoxic T-lymphocytes, cytotoxic T/NK lymphocytes, CD16 positive and CD16 negative monocytes, and blasts cells with lineage orientation. A 5-colour flow cytometer, the Beckman Coulter FC500, was used for analysis. The gating strategy described by Faucher et al. (2007) was used. EDTA blood was analysed on 27 normal samples and 148 abnormal samples which demonstrated abnormal cell flags on the LH750. These samples included the presence of blast cells, immature granulocytes and abnormal lymphocytes. Results for most cell populations measured by the flow cytometric differential compared well with both the LH750 automated differential and the manual reference method. Comparative results using Pearson correlation show that the automated LH750 differential produced r values of greater than 0.94 for neutrophils, lymphocytes and eosinophils. The manual reference method produced r values of greater than 0.89 for neutrophils, lymphocytes and eosinophils. Results for flow cytometric monocytes compared to the LH750 and manual differential gave an r value of 0.84 and 0.87 respectively. Results for basophils were significantly better when the flow cytometric method was compared to the LH750 rather than the manual method, r = 0.68 for flow cytometry versus LH750 and r = 0.43 for flow cytometry versus manual method. The value of the manual differential is diminished because of the low number of cells counted; the precision is not good for smaller cell populations (Hübl et al. 1995). Very good correlation of blast cells, r = 0.98 and immature granulocytes, r = 0.92 was seen between the manual and flow cytometric method. The flow cytometric differential is superior to the microscopic method since it is objective and due to the higher number of cells counted, it can detect subpopulations of cells that are present in smaller number with greater statistical and interpretive confidence. More importantly, it recognises and quantitates morphologically abnormal cells such as reactive lymphocytes, inflammatory monocytes and the lineage of blast cells. However, the examination of blood cell morphology by microscopy still has an important role in the diagnosis of diseases. Disclosures: No relevant conflicts of interest to declare.


2015 ◽  
Vol 69 (7) ◽  
pp. 637-642 ◽  
Author(s):  
Antoinette Mifsud ◽  
Daphne Peelen ◽  
Patricia Brincat ◽  
Sylvana Abela ◽  
Neville Debattista ◽  
...  

AimsThe aim of this study was to check the effect of Triton X-100 on various, commonly used haematology test parameters.MethodsAnonymised blood samples were treated with 10 µL of 10% Triton X-100 per 1 mL of blood. Treated and untreated samples were tested in parallel for blood film morphology, complete blood counts (CBCs), flow cytometry, blood grouping and antibody screens. Samples were also taken in 3.2% citrate tubes for coagulation test analyses.ResultsStatistical differences were noted in all CBC parameters apart from the mean cell volume, eosinophil and basophil counts. Platelet counts were significantly different with an apparent rise after the addition of Triton X-100. Samples were noted to have a high red cell fragmentation index. Immunological platelet counting methods using flow cytometry and fluorescent methods showed no significant differences and gave reliable results. Neither flow cytometry for T-cell subsets nor blood grouping/antibody screens were affected by Triton X-100. However, coagulation samples were severely haemolysed prohibiting analysis.ConclusionsWe have demonstrated that the addition of Triton X-100 to haematology blood samples impacts mainly on platelet counts and coagulation studies due to haemolysis. The platelet count is spuriously raised probably due to the presence of red cell fragments. The latter can be circumvented by the use of immunological platelet counting technology.


2005 ◽  
Vol 12 (1) ◽  
pp. 224-227 ◽  
Author(s):  
Godwin E. Imade ◽  
Bitrus Badung ◽  
Sunday Pam ◽  
Oche Agbaji ◽  
Daniel Egah ◽  
...  

ABSTRACT We compared two techniques for CD4 T-lymphocyte counting: flow cytometry (Cyflow) and magnetic beads (Dynabead). Similar results with good correlation were obtained from the 40 adult blood samples counted (P = 0.057, r = 0.93). The Cyflow technique is more precise and cost-effective than the Dynabead method ($3 to $5 versus $12 to $22 per test, respectively), since as many as 200 samples can be measured per day.


Cancers ◽  
2021 ◽  
Vol 13 (18) ◽  
pp. 4690
Author(s):  
Brian Hayes ◽  
Lauren Brady ◽  
Gráinne Sheill ◽  
Anne-Marie Baird ◽  
Emer Guinan ◽  
...  

Interactions between circulating tumour cells (CTCs) and platelets are thought to inhibit natural killer(NK)-cell-induced lysis. We attempted to correlate CTC numbers in men with advanced prostate cancer with platelet counts and circulating lymphocyte numbers. Sixty-one ExPeCT trial participants, divided into overweight/obese and normal weight groups on the basis of a BMI ≥ 25 or <25, were randomized to participate or not in a six-month exercise programme. Blood samples at randomization, and at three and six months, were subjected to ScreenCell filtration, circulating platelet counts were obtained, and flow cytometry was performed on a subset of samples (n = 29). CTC count positively correlated with absolute total lymphocyte count (r2 = 0.1709, p = 0.0258) and NK-cell count (r2 = 0.49, p < 0.0001). There was also a positive correlation between platelet count and CTC count (r2 = 0.094, p = 0.0001). Correlation was also demonstrated within the overweight/obese group (n = 123, p < 0.0001), the non-exercise group (n = 79, p = 0.001) and blood draw samples lacking platelet cloaking (n = 128, p < 0.0001). By flow cytometry, blood samples from the exercise group (n = 15) had a higher proportion of CD3+ T-lymphocytes (p = 0.0003) and lower proportions of B-lymphocytes (p = 0.0264) and NK-cells (p = 0.015) than the non-exercise group (n = 14). These findings suggest that CTCs engage in complex interactions with the coagulation cascade and innate immune system during intravascular transit, and they present an attractive target for directed therapy at a vulnerable stage in metastasis.


Blood ◽  
2010 ◽  
Vol 116 (21) ◽  
pp. 4726-4726
Author(s):  
Guat Bee Tan ◽  
Christina Sum ◽  
Ponnudurai Kuperan

Abstract Abstract 4726 Automated blood cell counters provide a leucocyte count and five-part (some 6-part) leucocyte differential, however haematology instrument differentials provide only limited information on cell morphology using abnormal cell flags and are often unable to reliably classify abnormal and immature cells. There are also limitations with the standard microscopic differential, identification of cells is subjective and there is significant inter and intra observer variation (Koepke et al. 1985). It is also subject to significant statistical variance (Rumke 1985). There have recently been reports of using monoclonal antibody cocktails for an extended leucocyte differential by flow cytometry. The International Council for Standardization in Haematology has set up a group to prepare an international reference method for an extended flow differential; this is intended to replace the current reference manual microscopic 2 × 200 cell count (CLSI H20-A2). Currently, there are several different protocols in use for leucocyte differential using different monoclonal antibodies and gating strategies. The aim of this study was to compare the differential results from a protocol from Singapore (SGP) with published flow cytometric protocols for the leucocyte differential from France (Faucher et al. 2007, Roussel et al. 2010) and to the automated count from the Beckman Coulter LH750 analyser and the current reference microscopic method. The French flow cytometric method uses 6 antibodies and allows detection of all white blood cells, mature neutrophils, total lymphocytes, total monocytes, eosinophils, basophils, immature granulocytes, B lymphocytes, non-cytotoxic T-lymphocytes, cytotoxic T/NK lymphocytes, CD16 positive and CD16 negative monocytes, and blasts cells with lineage orientation. The SGP method uses 8 antibodies (CD3, CD34, CD117, CD45, CD13, CD20, CD16, CD56), premixed in single tube. It detects mature neutrophils, total lymphocytes, total monocytes, eosinophils, basophils, CD16 positive and CD16 negative monocytes, T-lymphocytes, B-Lymphocytes, NK-cells, immature granulocytes and blasts. A 5-colour flow cytometer, the Beckman Coulter FC500, was used in this study. EDTA blood was analysed on 27 normal and 148 abnormal samples, either with complete blood count values outside the reference range or which demonstrated abnormal cell flags on the LH750. These samples included blast cells, immature granulocytes and abnormal lymphocytes. Results for most cell populations measured by the SGP flow differential compared well with the LH750, the manual reference method and French protocol. Comparative results using Pearson correlation are presented in Table 1. For the SGP protocol, correlation with the LH750 and with the manual differential was good for neutrophils, lymphocytes, monocytes and eosinophils. Excellent correlation was observed for all cells apart from basophils when the two flow methods were compared to each other. There was no correlation for basophils between the SGP flow method and the manual method. Similarly, there was no correlation between SGP flow method and LH750 nor between both flow methods. This is not surprising as basophils are usually present in very low numbers. Hence, without a positive marker for basophils in the flow cytometric panel correlation may depend on the type of samples used for the evaluation. Very good correlation of blast cells, r=0.99 and immature granulocytes, r=0.88 was seen between the manual and the SGP method. When comparing the flow methods to each other correlation for blast cells shows an r value of 0.96 and immature granulocytes 0.97. Our study shows that this flow cytometric method performs well with both normal and abnormal patient samples. A differential using monoclonal antibodies for immunological recognition of cells provides more information than either the manual or automated differential. In addition to the detection of the common cell populations, blast cells, immature granulocytes, subpopulations of lymphocytes and inflammatory monocytes are enumerated. Disclosures: No relevant conflicts of interest to declare.


1989 ◽  
Vol 62 (04) ◽  
pp. 1107-1111 ◽  
Author(s):  
Hugo C Castro-Faria-Neto ◽  
Patricia T Bozza ◽  
Marco A Martins ◽  
Paulo M F L Dias ◽  
Patricia M R Silva ◽  
...  

SummaryThe injection of PAP (6 μg/kg, i. v.) induced, in rats, haemoconcentration accompanied by an increase in the platelet number, as attested by the counts of platelets in blood samples diluted in formalin-free EDTA solution. This increase was significant at 15 min, peaked from 1 to 4 h and returned to basal levels 24 h after the lipid administration. The release of platelets induced by PAP was inhibited dose-dependently by specific PAP receptor antagonists such as WEB 2086 (0.5-2 mg/kg), BN 52021 and 48740 RP (5-25 mg/kg). Furthermore, platelet mobilization was clearly impaired in splenectomized animals stimulated by PAP, whereas thrombocytopenia and haemoconcentration by the same stimulus were intact. It was also noted that a second injection of PAP, 24 h after the initial stimulation with the lipid, failed to induce an increase in platelet counts, indicating autodesensitization. Desensitization to PAP or pretreatment with PAP antagonists clearly prevented the increase in the platelet counts after stimulation by adrenaline (15 μg/kg). These findings suggest that, in rats, PAP can induce release of platelets by a spleen-dependent mechanism and that this lipid may be relevant to the thrombocytosis triggered by adrenaline.


Sign in / Sign up

Export Citation Format

Share Document