scholarly journals Design and Construction of an Optical TEM Specimen Holder

2021 ◽  
Vol 27 (S1) ◽  
pp. 2310-2312
Author(s):  
Joel Martis ◽  
Ze Zhang ◽  
Haokun Li ◽  
Arun Majumdar ◽  
Roy Kim ◽  
...  
Author(s):  
Russell L. Steere ◽  
Michael Moseley

A redesigned specimen holder and cap have made possible the freeze-etching of both fracture surfaces of a frozen fractured specimen. In principal, the procedure involves freezing a specimen between two specimen holders (as shown in A, Fig. 1, and the left side of Fig. 2). The aluminum specimen holders and brass cap are constructed so that the upper specimen holder can be forced loose, turned over, and pressed down firmly against the specimen stage to a position represented by B, Fig. 1, and the right side of Fig. 2.


Author(s):  
Louis T. Germinario

A liquid nitrogen stage has been developed for the JEOL JEM-100B electron microscope equipped with a scanning attachment. The design is a modification of the standard JEM-100B SEM specimen holder with specimen cooling to any temperatures In the range ~ 55°K to room temperature. Since the specimen plane is maintained at the ‘high resolution’ focal position of the objective lens and ‘bumping’ and thermal drift la minimized by supercooling the liquid nitrogen, the high resolution capability of the microscope is maintained (Fig.4).


Author(s):  
David C. Joy ◽  
Dennis M. Maher

High-resolution images of the surface topography of solid specimens can be obtained using the low-loss technique of Wells. If the specimen is placed inside a lens of the condenser/objective type, then it has been shown that the lens itself can be used to collect and filter the low-loss electrons. Since the probeforming lenses in TEM instruments fitted with scanning attachments are of this type, low-loss imaging should be possible.High-resolution, low-loss images have been obtained in a JEOL JEM 100B fitted with a scanning attachment and a thermal, fieldemission gun. No modifications were made to the instrument, but a wedge-shaped, specimen holder was made to fit the side-entry, goniometer stage. Thus the specimen is oriented initially at a glancing angle of about 30° to the beam direction. The instrument is set up in the conventional manner for STEM operation with all the lenses, including the projector, excited.


Author(s):  
Tokio Nei ◽  
Haruo Yotsumoto ◽  
Yoichi Hasegawa ◽  
Yuji Nagasawa

In order to observe biological specimens in their native state, that is, still containing their water content, various methods of specimen preparation have been used, the principal two of which are the chamber method and the freeze method.Using its recently developed cold stage for installation in the pre-evacuation chamber of a scanning electron microscope, we have succeeded in directly observing a biological specimen in its frozen state without the need for such conventional specimen preparation techniques as drying and metallic vacuum evaporation. (Echlin, too, has reported on the observation of surface structures using the same freeze method.)In the experiment referred to herein, a small sliced specimen was place in the specimen holder. After it was rapidly frozen by freon cooled with liquid nitrogen, it was inserted into the cold stage of the specimen chamber.


Author(s):  
William P. Wergin ◽  
Eric F. Erbe ◽  
Eugene L. Vigil

Investigators have long realized the potential advantages of using a low temperature (LT) stage to examine fresh, frozen specimens in a scanning electron microscope (SEM). However, long working distances (W.D.), thick sputter coatings and surface contamination have prevented LTSEM from achieving results comparable to those from TEM freeze etch. To improve results, we recently modified techniques that involve a Hitachi S570 SEM, an Emscope SP2000 Sputter Cryo System and a Denton freeze etch unit. Because investigators have frequently utilized the fractured E face of the plasmalemma of yeast, this tissue was selected as a standard for comparison in the present study.In place of a standard specimen holder, a modified rivet was used to achieve a shorter W.D. (1 to -2 mm) and to gain access to the upper detector. However, the additional height afforded by the rivet, precluded use of the standard shroud on the Emscope specimen transfer device. Consequently, the sample became heavily contaminated (Fig. 1). A removable shroud was devised and used to reduce contamination (Fig. 2), but the specimen lacked clean fractured edges. This result suggested that low vacuum sputter coating was also limiting resolution.


Author(s):  
Earl J. Kirkland ◽  
Robert J. Keyse

An ultra-high resolution pole piece with a coefficient of spherical aberration Cs=0.7mm. was previously designed for a Vacuum Generators HB-501A Scanning Transmission Electron Microscope (STEM). This lens was used to produce bright field (BF) and annular dark field (ADF) images of (111) silicon with a lattice spacing of 1.92 Å. In this microscope the specimen must be loaded into the lens through the top bore (or exit bore, electrons traveling from the bottom to the top). Thus the top bore must be rather large to accommodate the specimen holder. Unfortunately, a large bore is not ideal for producing low aberrations. The old lens was thus highly asymmetrical, with an upper bore of 8.0mm. Even with this large upper bore it has not been possible to produce a tilting stage, which hampers high resolution microscopy.


Author(s):  
W.K. Lo ◽  
J.C.H. Spence

An improved design for a combination Scanning Tunnelling Microscope/TEM specimen holder is presented. It is based on earlier versions which have been used to test the usefulness of such a device. As with the earlier versions, this holder is meant to replace the standard double-tilt specimen holder of an unmodified Philips 400T TEM. It allows the sample to be imaged simultaneously by both the STM and the TEM when the TEM is operated in the reflection mode (see figure 1).The resolution of a STM is determined by its tip radii as well as its stability. This places strict limitations on the mechanical stability of the tip with respect to the sample. In this STM the piezoelectric tube scanner is rigidly mounted inside the endcap of the STM holder. The tip coarse approach to the sample (z-direction) is provided by an Inchworm which is located outside the TEM vacuum.


Author(s):  
Edward A Kenik

Segregation of solute atoms to grain boundaries, dislocations, and other extended defects can occur under thermal equilibrium or non-equilibrium conditions, such as quenching, irradiation, or precipitation. Generally, equilibrium segregation is narrow (near monolayer coverage at planar defects), whereas non-equilibrium segregation exhibits profiles of larger spatial extent, associated with diffusion of point defects or solute atoms. Analytical electron microscopy provides tools both to measure the segregation and to characterize the defect at which the segregation occurs. This is especially true of instruments that can achieve fine (<2 nm width), high current probes and as such, provide high spatial resolution analysis and characterization capability. Analysis was performed in a Philips EM400T/FEG operated in the scanning transmission mode with a probe diameter of <2 nm (FWTM). The instrument is equipped with EDAX 9100/70 energy dispersive X-ray spectrometry (EDXS) and Gatan 666 parallel detection electron energy loss spectrometry (PEELS) systems. A double-tilt, liquid-nitrogen-cooled specimen holder was employed for microanalysis in order to minimize contamination under the focussed spot.


Author(s):  
E. B. Steel

High Purity Germanium (HPGe) x-ray detectors are now commercially available for the analytical electron microscope (AEM). The detectors have superior efficiency at high x-ray energies and superior resolution compared to traditional lithium-drifted silicon [Si(Li)] detectors. However, just as for the Si(Li), the use of the HPGe detectors requires the determination of sensitivity factors for the quantitative chemical analysis of specimens in the AEM. Detector performance, including incomplete charge, resolution, and durability has been compared to a first generation detector. Sensitivity factors for many elements with atomic numbers 10 through 92 have been determined at 100, 200, and 300 keV. This data is compared to Si(Li) detector sensitivity factors.The overall sensitivity and utility of high energy K-lines are reviewed and discussed. Many instruments have one or more high energy K-line backgrounds that will affect specific analytes. One detector-instrument-specimen holder combination had a consistent Pb K-line background while another had a W K-line background.


Author(s):  
K. Fukushima ◽  
T. Kaneyama ◽  
F. Hosokawa ◽  
H. Tsuno ◽  
T. Honda ◽  
...  

Recently, in the materials science field, the ultrahigh resolution analytical electron microscope (UHRAEM) has become a very important instrument to study extremely fine areas of the specimen. The requirements related to the performance of the UHRAEM are becoming gradually severer. Some basic characteristic features required of an objective lens are as follows, and the practical performance of the UHRAEM should be judged by totally evaluating them.1) Ultrahigh resolution to resolve ultrafine structure by atomic-level observation.2) Nanometer probe analysis to analyse the constituent elements in nm-areas of the specimen.3) Better performance of x-ray detection for EDS analysis, that is, higher take-off angle and larger detection solid angle.4) Higher specimen tilting angle to adjust the specimen orientation.To attain these requirements simultaneously, the objective lens polepiece must have smaller spherical and chromatic aberration coefficients and must keep enough open space around the specimen holder in it.


Sign in / Sign up

Export Citation Format

Share Document