The Reliability of Point-of-Care (POC) Prothrombin Time Testing - A Comparison of CoaguChek XS® INR Measurements with Hospital Laboratory Monitoring.

Blood ◽  
2007 ◽  
Vol 110 (11) ◽  
pp. 973-973 ◽  
Author(s):  
Fiona Ryan ◽  
Susan O’Shea ◽  
Thomas L. Ortel ◽  
Stephen Byrne

Abstract Background The development of Point of Care (POC) testing devices enables patients to test their own International Normalised Ratio (INR) at home, and has lead to the development of patient self- testing (PST) programmes. This facilitates more frequent testing of anticoagulation levels which in turn improves time in therapeutic range (TTR) and reduces bleeding and thrombotic complications in patients on oral anticoagulation therapy (OAT).1 External quality control of POC devices is essential to ensure that results are robust. One approach is to collect a venous sample at the same time as POC testing, which is analysed using an appropriate hospital laboratory analyzer. It has been recommended that INR results should be within 0.5 of each other when measured by each method2. However, previous studies have shown that when compared with clinical laboratory values, POC devices may show statistically significant differences between the INR values.3 Aim The aim of this study is primarily to compare TTR and adverse events between two different approaches to OAT management and more specifically in this presentation to examine the intra patient variability of INR measurements performed by two different techniques. Methods 160 patients routinely attending the anticoagulant clinic of Cork University Hospital (CUH) were enrolled to take part in a prospective randomized controlled, cross-over design trial of supervised PST of warfarin therapy using an internet based expert system (Coagcare®, ZyCare, Inc) and POC testing device (CoaguChek XS®, Roche, UK) vs. rountine medical care in CUH. During the 6 months PST arm of the study a venous blood sample was taken from each patient for laboratory determination of the INR on 3 occasions; at enrolment, 2 months and 4 months. This measurement was then compared with the INR measured on the Coaguchek XS® machine using capillary blood from the patient’s fingertip. Results 88 patients have being enrolled in the PST arm of the study to date and 454 POC measurements have been compared with laboratory values. Good correlation has been shown between the two methods of INR determination (r2 = 0.8314), however, statistically significant differences did occur (p < 0.005). A Bland and Altman plot showed good agreement between the dual measurements at values between 2.0–3.5 INR units. However there seemed to be increasing disagreement between the two methods as the INR rose above 3.5. Four patients were excluded from the study because the differences in the INR measurements were greater than 0.5 INR units. Conclusion POC testing devices for OAT are reliable and safe. External quality assessment should be inherent to any programme involving these devices. Clinicians involved in the use of POC anticoagulation monitoring should be aware that POC and laboratory values may differ and intervene if there is not adequate correlation between the two INR measurements.

2008 ◽  
Vol 99 (06) ◽  
pp. 1097-1103 ◽  
Author(s):  
Karina Black ◽  
Mary Massicotte ◽  
Michelle Bauman ◽  
Stefan Kuhle ◽  
Susan Howlett-Clyne ◽  
...  

SummaryPoint-of-care INR (POC INR) meters can provide a safe and effective method for monitoring oral vitamin K antagonists (VKAs) in children. Stollery Children’s Hospital has a large POC INR meter loan program for children requiring oral VKAs. Our protocol requires that POC INR results be compared to the standard laboratory INR for each child on several consecutive tests to ensure accuracy of CoaguChek XS® (Roche Diagnostics, Basel Switzerland) meter. It was the objective of the study to determine the accuracy of the CoaguChek XS by comparing whole blood INR results from the CoaguChek XS to plasma INR results from the standard laboratory in children. POC INR meter validations were performed on plasma samples from two time points from 62 children receiving warfarin by drawing a venous blood sample for laboratory prothrombin (PT)-INR measurements and simultaneous INR determinations using the POC-INR meter. Agreement between CoaguChek XS INR and laboratory INR was assessed using Bland-Altman plots. Bland-Altman's 95% limits of agreement were 0.11 (-0.20; 0.42) and 0.13 (-0.22; 0.48) at the two time points, respectively. In conclusion, the CoaguChek XS meter appraisal generates an accurate and precise INR measure in children when compared to laboratory INR test results.


2015 ◽  
Vol 133 (6) ◽  
pp. 460-464 ◽  
Author(s):  
Aurélie Affret ◽  
Luiz Henrique Maciel Griz ◽  
Eduarda Ângela Pessoa Cesse ◽  
Yuri da Silva Specht ◽  
Eduardo Maia Freese de Carvalho ◽  
...  

CONTEXT AND OBJECTIVE: To monitor glycemic control in diabetic patients, regular measurement of glycated hemoglobin (HbA1c) is recommended, but this can be difficult in remote places without access to laboratories. Portable point-of-care testing devices can prove a useful alternative. Our study aimed to assess the performance of one of them: A1CNow+, from Bayer. DESIGN AND SETTING: Cross-sectional accuracy study conducted at a university hospital in Brazil. METHODS: We made three successive measurements of capillary HbA1c using the A1CNow+ in 55 diabetic volunteers, while the same measurement was made on venous blood using the hospital reference method (Vitros 5,1 FS). We used the Bland-Altman graphical method to assess the A1CNow+ in relation to the Vitros 5,1 FS method. We also evaluated clinical usefulness by calculating the sensitivity and specificity of A1CNow+ for detecting patients with HbA1c lower than 7%, which is the usual limit for good glycemic control. RESULTS: The coefficient of variation between repeat testing for the A1CNow+ was 3.6%. The mean difference between A1CNow+ and Vitros 5,1 FS was +0.67% (95% confidence interval, CI: +0.52 to +0.81). The agreement limits of our Bland-Altman graph were -0.45 (95% CI: -0.71 to -0.19) and +1.82 (95% CI: +1.52 to +2.05). The sensitivity and specificity in relation to the 7% limit were respectively 100% and 67.7%. CONCLUSIONS: Although the A1CNow+ had good sensitivity, its accuracy was insufficient for use as a replacement for laboratory measurements of HbA1c, for glycemic control monitoring in diabetic patients.


2007 ◽  
Vol 97 (05) ◽  
pp. 586-861 ◽  
Author(s):  
Ingrid Leichsenring ◽  
Winfried Plesch ◽  
Volker Unkrig ◽  
Steve Kitchen ◽  
Dianne Kitchen ◽  
...  

SummaryThe new CoaguChek XS system is designed for use in patient selftesting. It is the successor of the current CoaguChek S system. The detection principle is based on the amperometric measurement of the thrombin activity initiated by starting the coagulation cascade using a human recombinant thromboplastin. This study was performed to assign the International SEnsitivity Index (ISI) to the new test according to the WHO guidelines for thromboplastins and plasmas used to control anticoagulant therapy, and to establish the measuring range of the new system. At four study sites a total of 90 samples of normal donors and 291 samples of warfarin-, phenprocoumon- or acenocoumarol-treated patients were included in the study.The ISI value of the new test was assigned against the human recombinant reference thromboplastin rTF/95 at each site using the samples from stabilized patients in the International Normalized Ratio (INR) range between 1.5 and 4.5 only. The new point-of-care system’s measuring range between 0.8 and 8 INR was calibrated against the mean INR of rTF/95 and AD149 using polynomial regression. ISIs were (CV of the slope): Site 1: ISI 0.99 (1.1%); Site 2: ISI 1.02 (2.0%); Site 3: ISI 1.03 (1.1%); Site 4: ISI 1.00 (1.4%). All regression lines calculated from patient-only data pass through the normal donor data points.All CVs of the slopes of the orthogonal regression lines are well below 3%,thus fulfilling the requirements of the WHO guidelines. The mean ISI for the new CoaguChek XS PT Test is 1.01.


2020 ◽  
Vol 120 (04) ◽  
pp. 687-691
Author(s):  
Majed A. Refaai ◽  
Vinay Shah ◽  
Ronald Fernando

Abstract Introduction There are limited publications about the microINR point-of-care (POC) system (iLine Microsystems). The current microINR POC system was compared with the ACL TOP 500 laboratory analyzer (Instrumentation Laboratory) and with the CoaguChek XS POC system (Roche Diagnostics). Methods This study was performed at three United States medical centers. Sixty-eight nonanticoagulated normal donors and 245 warfarin anticoagulated patients were included. Testing was performed in duplicate using capillary blood samples for the POC systems and venous blood samples for the laboratory testing. Accuracy and imprecision were assessed. Results Comparing microINR to ACL revealed a correlation coefficient (r) of 0.973, a slope of 1.00 (95% confidence interval [CI], 0.97–1.03), and an intercept of 0.08 (95% CI, 0.04–0.15). When compared with the CoaguChek XS, r was 0.977 with a slope of 0.92 (95% CI, 0.89–0.94) and an intercept of 0.15 (95% CI, 0.08–0.19). Predicted bias values at international normalized ratio (INR) 2.0, 3.5, and 4.5 were ≤ 5% against both references. Agreement with ACL was 97, 95, and 100% for the INR ranges of < 2.0 ± 0.40, 2.0 to 4.5 ± 20%, and ≥ 4.5 ± 25%, respectively. Agreement for the combined INR ranges was 96% against ACL and > 96% against the CoaguChek XS. The coefficient of variation of the microINR was 5.03% for INR < 2.0 and 4.68% for the therapeutic INR range 2.0 to 3.5. Conclusion The microINR results demonstrate adequate imprecision and accuracy to both ACL and CoaguChek XS. This indicates that monitoring INR by this microINR POC system is reliable and acceptable for the management of warfarin therapy.


2009 ◽  
Vol 3 (3) ◽  
pp. 418-423 ◽  
Author(s):  
Erna Lenters-Westra ◽  
Robbert J. Slingerland

Background: Point-of-care instruments for the measurement of hemoglobin A1c (HbA1c) may improve the glycemic control of people with diabetes by providing a rapid result if the performance of the instruments used is acceptable. A 0.5% HbA1c difference between successive results is considered a clinically relevant change. With this in mind, the In2it from Bio-Rad and the DCA Vantage from Siemens were evaluated according to Clinical and Laboratory Standards Institute (CLSI) protocols. Methods: The CLSI protocols EP-5 and EP-9 were applied to investigate precision, accuracy, and bias. The bias was compared with three certified secondary reference measurement procedures. Differences between capillary and venous blood were investigated by an end-user group consisting of nurse practitioners at a diabetes care center. Results: At HbA1c levels of 5.1 and 11.2%, total coefficients of variation (CV) for the In2it were 4.9 and 3.3%, respectively, and for the DCA Vantage were 1.7 to 1.8% and 3.7 to 5.5% depending on the lot number of the cartridges. Method comparisons showed significant lot number-dependent results for the In2it and the DCA Vantage compared with the three reference methods. No overall difference was observed between capillary and venous blood for both methods. Conclusion: Performance results of the In2it and the DCA Vantage showed variable and lot number-dependent results. To maintain the interlaboratory CV of 5% for HbA1c, the Clinical Laboratory Improvement Amendments rules for waived point-of-care instruments should be revised. An obligation for participating in external quality schemes and taking adequate action should be considered for POC instruments that perform poorly.


Author(s):  
W Maule

Point-of-care testing (POCT) in haemostasis has shown a steady state increase in the range of tests available with an exponential increase in the number of tests that are performed annually. The most commonly performed test remains the out of hospital measurement of the international normalised ratio (INR). The INR is used to monitor oral anticoagulation therapy (OAT) with vitamin K antagonists. Such drugs include some of the following: warfarin (Coumadin®), phenprocoumon (Marcumar®) and acenocoumarol (Sintrom®, also sold as other brand names). Contrary to laboratory testing, POCT is often performed by general practitioners (GPs), nurses, pharmacists, patients and other healthcare professionals. In many cases these individuals do not have access to the same support mechanisms that exist in the clinical laboratory regarding strict quality control programmes. When an INR test is performed in the clinical laboratory, it has to be accredited, monitored and inspected. Accreditation is also advisable and part of these requirements is to be involved in external quality assessment (EQA). In the case of POCT haemostasis assays, EQA programmes do exist, however these programmes are still relatively few in number. Over the last decade there has been a significant increase in the scope of POCT, particularly in the field of haemostasis. This brief review summarises POCT in haemostasis, highlighting some of its benefits, challenges and future perspectives, especially with regard to the measurement of the INR.


2017 ◽  
Vol 2017 ◽  
pp. 1-5 ◽  
Author(s):  
I. P. L. Houben ◽  
C. J. L. Y. van Berlo ◽  
O. Bekers ◽  
E. C. Nijssen ◽  
M. B. I. Lobbes ◽  
...  

Purpose. To evaluate whether a handheld point-of-care (POC) device is able to predict and discriminate patients at potential risk of contrast-induced nephropathy (CIN) prior to iodine-based contrast media delivery. Methods and Materials. Between December 2014 and June 2016, women undergoing contrast-enhanced spectral mammography (CESM) with an iodine-based contrast agent were asked to have their risk of CIN assessed by a dedicated POC device (StatSensor CREAT) and a risk factor questionnaire based on national guidelines. Prior to contrast injection, a venous blood sample was drawn to compare the results of POC with regular laboratory testing. Results. A total of 351 patients were included; 344 were finally categorized as low risk patients by blood creatinine evaluation. Seven patients had a eGFR below 60 ml/min/1.73 m2, necessitating additional preparation prior to contrast delivery. The POC device failed to categorize six out of seven patients (86%), leading to (at that stage) unwanted contrast administration. Two patients subsequently developed CIN after 2–5 days, which was self-limiting after 30 days. Conclusion. The POC device tested was not able to reliably assess impairment of renal function in our patient cohort undergoing CESM. Consequently, we still consider classic clinical laboratory testing preferable in patients at potential risk for developing CIN.


2009 ◽  
Vol 18 (3) ◽  
pp. 224-230 ◽  
Author(s):  
Anna Shearer ◽  
Marilyn Boehmer ◽  
Melanie Closs ◽  
Rosalina Dela Rosa ◽  
Jean Hamilton ◽  
...  

Background Blood from central venous or arterial catheters as well as from fingersticks is often used for point-of-care glucose testing.Objectives To compare glucose values obtained with a point-of-care device for catheter and fingerstick blood samples with values obtained by clinical laboratory analysis of a catheter blood sample.Methods A method-comparison design was used. Point-of-care values for fingerstick and catheter blood samples were compared with laboratory values for blood from catheters in a convenience sample of 63 critically ill patients. Device bias and precision were calculated and graphed according to the Bland-Altman method, and a t test was used to determine differences in glucose values for the 2 methods.Results Laboratory glucose values for blood from a catheter sample differed significantly from point-of-care values for blood from a fingerstick (t1,61 = 5.01; P &lt; .001) and from a catheter (t1,61 = 3.91; P &lt; .001). Bias and precision for the point-of-care device was large (fingerstick, 8.7 and 13.7; catheter, 7.0 and 14.0); 20% of the values differed from the laboratory glucose values by more than 20 mg/dL. Point-of-care glucose values for fingerstick and catheter samples did not differ (P = .41).Conclusions Glucose values obtained with a point-of-care device differ significantly from those obtained by laboratory analysis. The magnitude of these differences calls into question the widespread practice of using point-of-care glucose testing to guide insulin titration for tight glucose control. Errors in dosing could easily be made because of the large bias and precision associated with a point-of-care device.


2019 ◽  
Vol 34 (05) ◽  
pp. 506-509
Author(s):  
Jessica Topping ◽  
Matthew Reardon ◽  
Jake Coleman ◽  
Brian Hunter ◽  
Haruka Shojima-Perera ◽  
...  

AbstractBackground:Blood glucose level (BGL) is routinely assessed by paramedics in the out-of-hospital setting. Most commonly, BGL is measured using a blood sample of capillary origin analyzed by a hand-held, point-of-care glucometer. In some clinical circumstances, the capillary sample may be replaced by blood of venous origin. Given most point-of-care glucometers are engineered to analyze capillary blood samples, the use of venous blood instead of capillary may lead to inaccurate or misleading measurements.Hypothesis/Problem:The aim of this prospective study was to compare mean difference in BGL between venous and capillary blood from healthy volunteers when measured using a capillary-based, hand-held, point-of-care glucometer.Methods:Using a prospective observational comparison design, 36 healthy participants provided paired samples of blood, one venous and the other capillary, taken near simultaneously. The BGL values were similar between the two groups. The capillary group had a range of 4.3mmol/l, with the lowest value being 4.4mmol/l and 8.7mmol/l the highest. The venous group had a range of 2.7mmol/l, with the lowest value being 4.1mmol/l and 7.0mmol/l the highest.For the primary research question, the mean BGL for the venous sample group was 5.3mmol/l (SD = 0.6), compared to 5.6mmol/l (SD = 0.8) for the capillary group. This represented a statistically significant difference of 0.3mmol/l (P = .04), but it did not reach the a priori established point of clinical significance (1.0mmol/l). Pearson’s correlation coefficient for capillary versus venous indicated moderate correlation (r = 0.42).Conclusion:In healthy, non-fasted people in a non-clinical setting, a statistically significant, but not clinically significant, difference was found between venous- and capillary-derived BGL when measured using a point-of-care, capillary-based glucometer. Correlation between the two was moderate. In this context, using venous samples in a capillary-based glucometer is reasonable providing the venous sample can be gathered without exposure of the clinician to risk of needle-stick injury. In clinical settings where physiological derangement or acute illness is present, capillary sampling would remain the optimal approach.


1990 ◽  
Vol 36 (2) ◽  
pp. 255-260 ◽  
Author(s):  
P S Bachorik ◽  
R Rock ◽  
T Cloey ◽  
E Treciak ◽  
D Becker ◽  
...  

Abstract We measured cholesterol in capillary blood samples from 9683 volunteers over a four-day on-site community screening program, using the "Reflotron" desk-top analyzer (Boehringer-Mannheim Diagnostics, Indianapolis, IN). We also measured cholesterol in venous blood samples from 3% of those screened (a) with the Reflotron at the screening sites, (b) in a qualified hospital clinical laboratory, and (c) in a Centers for Disease Control standardized lipoprotein research laboratory. The sensitivity (and specificity) of the Reflotron measurements, with use of the lipoprotein laboratory measurements as the point of reference, was 0.95 (0.73) in capillary blood samples and 0.88 (0.93) in venous blood samples, compared with 0.99 (0.87) in the hospital clinical laboratory. The Reflotron measurements correlated less well with the lipoprotein laboratory values in both venous blood (r = 0.91) and capillary blood (r = 0.89) samples than did the clinical laboratory values (r greater than 0.99). Furthermore, the capillary blood measurements averaged 7% higher than venous measurements when both kinds of samples were analyzed in the Reflotron.


Sign in / Sign up

Export Citation Format

Share Document