Heparin and citrate additive carryover during blood collection

2019 ◽  
Vol 57 (12) ◽  
pp. 1888-1896 ◽  
Author(s):  
Martin H. Keppel ◽  
Simon Auer ◽  
Giuseppe Lippi ◽  
Alexander von Meyer ◽  
Michael Cornes ◽  
...  

Abstract Background Published evidence on the risk of additive carryover during phlebotomy remains elusive. We aimed to assess potential carryover of citrated and heparinized blood and the relative volume needed to bias clinical chemistry and coagulation tests. Methods We simulated standardized phlebotomies to quantify the risk of carryover of citrate and heparin additives in distilled water, using sodium and lithium as surrogates. We also investigated the effects of contamination of heparinized blood samples with increasing volumes of citrated blood and pure citrate on measurements of sodium, potassium, chloride, magnesium, total and ionized calcium and phosphate. Likewise, we studied the effects of contamination of citrated blood samples with increasing volumes of heparinized blood on heparin (anti-Xa) activity, lithium, activated partial thromboplastin time (APTT), prothrombin time (PT) and thrombin time (TT). We interpreted these results based on measurement deviations beyond analytical, biological and clinical significance. Results Standardized phlebotomy simulations revealed no significant differences in concentration of surrogate markers. Clinically significant alterations were observed after contamination of heparinized blood samples with volumes of citrated blood beyond 5–50 μL for ionized calcium and beyond 100–1000 μL for sodium, chloride and total calcium. Investigations of pure citrate carryover revealed similar results at somewhat lower volumes. Heparinized blood carryover showed clinically significant interference of coagulation testing at volumes beyond 5–100 μL. Conclusions Our results suggest that during a standardized phlebotomy, heparin or citrate contamination is highly unlikely. However, smaller volumes are sufficient to severely alter test results when deviating from phlebotomy guidelines.

2013 ◽  
Vol 137 (11) ◽  
pp. 1641-1647 ◽  
Author(s):  
Panutsaya Tientadakul ◽  
Chulalak Kongkan ◽  
Wimol Chinswangwatanakul

Context.—Heparin contamination in blood samples may cause false prolongation of activated partial thromboplastin time (aPTT) and prothrombin time results. Polybrene can neutralize heparin, but it affects coagulation by itself. Objectives.——To determine the optimal concentration of polybrene to neutralize heparin, to determine the suitable sequence of reagents for the neutralization method performed on the analyzer at the same time as prothrombin time and aPTT testing, and to detect the heparin contamination in blood samples for coagulation tests in our hospital using this method. Design.—Various concentrations of heparin were added to 10 normal and 76 abnormal plasma samples to study the efficacy of polybrene. Two programs of reagent sequencing for aPTT with polybrene performed on the analyzer were tested. Samples suspected of heparin contamination according to our criteria were selected for neutralization during a 3-month period. Results.——The optimal final concentration of polybrene was 25 μg/mL. Polybrene should be added after the aPTT reagent to minimize its interference effect. Even though results of prothrombin time and aPTT after neutralization did not equal those before the spike of heparin, the differences might not be clinically significant. Eighty-one of 4921 samples (1.6%) were selected for aPTT with the neutralization method, and the detection rate of heparin contamination was 84% (68 of 81), giving an overall incidence of 1.4% (68 of 4921). Conclusions.—This method is inexpensive and can be performed rapidly with prothrombin time and aPTT on the automated analyzer, which makes it easy to practice with no need for extra plasma volumes.


Diagnostics ◽  
2021 ◽  
Vol 11 (6) ◽  
pp. 1019
Author(s):  
Kyungjin Hong ◽  
Gabriella Iacovetti ◽  
Ali Rahimian ◽  
Sean Hong ◽  
Jon Epperson ◽  
...  

Blood sample collection and rapid separation—critical preanalytical steps in clinical chemistry—can be challenging in decentralized collection settings. To address this gap, the Torq™ zero delay centrifuge system includes a lightweight, hand-portable centrifuge (ZDrive™) and a disc-shaped blood collection device (ZDisc™) enabling immediate sample centrifugation at the point of collection. Here, we report results from clinical validation studies comparing performance of the Torq System with a conventional plasma separation tube (PST). Blood specimens from 134 subjects were collected and processed across three independent sites to compare ZDisc and PST performance in the assessment of 14 analytes (K, Na, Cl, Ca, BUN, creatinine, AST, ALT, ALP, total bilirubin, albumin, total protein, cholesterol, and triglycerides). A 31-subject precision study was performed to evaluate reproducibility of plasma test results from ZDiscs, and plasma quality was assessed by measuring hemolysis and blood cells from 10 subject specimens. The ZDisc successfully collected and processed samples from 134 subjects. ZDisc results agreed with reference PSTs for all 14 analytes with mean % biases well below clinically significant levels. Results were reproducible across different operators and ZDisc production lots, and plasma blood cell counts and hemolysis levels fell well below clinical acceptance thresholds. ZDiscs produce plasma samples equivalent to reference PSTs. Results support the suitability of the Torq System for remotely collecting and processing blood samples in decentralized settings.


2022 ◽  
Vol 12 (1) ◽  
Author(s):  
Tongqing Chen ◽  
Duan Chen ◽  
Lu Chen ◽  
Zhengxu Chen ◽  
Baolong Wang ◽  
...  

AbstractTo evaluate the effects of fructose diphosphate (FDP) on routine coagulation tests in vitro, we added FDP into the mixed normal plasma to obtain the final concentration of 0, 1, 2, 3, 4, 5, 6, 10, 15, 20, 25, 30 and 35 mg/mL of drug. Prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen (FBG) and thrombin time (TT) of samples were analyzed with blood coagulation analyzers from four different manufacturers(Sysmex, Stago, SEKISUI and Werfen) and their corresponding reagents, respectively. Before the experiment, we also observed whether there were significant differences in coagulation test results of different lots of reagents produced by each manufacturer. At the same time as the four routine clotting tests, the Sysmex blood coagulation analyzer and its proprietary analysis software were used to detect the change of maximum platelet aggregation rate in platelet-rich plasma after adding FDP (0, 1, 2, 3, 4, 5 and 6 mg/mL). The results of PT, aPTT and TT showed a FDP (0–35 mg/mL) concentration-dependent increase and a FBG concentration-dependent decrease. The degree of change (increase or decrease) varied depending on the assay system, with PT and aPTT being more affected by the Sysmex blood coagulation testing instrument reagent system and less affected by CEKISUI, TT less affected by CEKISUI and more affected by Stago, and FBG less affected by Stago and more affected by Sysmex. The results of PT, aPTT and TT were statistically positively correlated with their FDP concentrations, while FBG was negatively correlated. The correlation coefficients between FDP and the coagulation testing systems of Sysmex, Stago, Werfen and SEKISUI were 0.975, 0.988, 0.967, 0.986 for PT, and 0.993, 0.989, 0.990 and 0.962 for aPTT, 0.994, 0.960, 0.977 and 0.982 for TT, − 0.990, − 0.983, − 0.989 and − 0.954 for FBG, respectively. Different concentrations of FDP (0, 1, 2, 3, 4, 5 and 6 mg/mL) had different effects on the maximum aggregation rate of platelet induced by the agonists of adenosine diphosphate (ADP, 5 µmol/L), arachidonic acid (Ara, 1 mmol/L), collagen (Col, 2.5 µg/mL) and epinephrine (Epi,10 µmol/L), but the overall downward trend was consistent, that is, with the increase of FDP concentration, the platelet aggregation rate decreased significantly. Our experimental study demonstrated a possible effect of FDP on the assays of coagulation and Platelet aggregation, which may arise because the drug interferes with the coagulation and platelet aggregation detection system, or it may affect our in vivo coagulation system and Platelet aggregation function, the real mechanism of which remains to be further verified and studied.


1994 ◽  
Vol 22 (6) ◽  
pp. 698-702 ◽  
Author(s):  
B. G. Carter ◽  
J. Tibballs ◽  
M. Hochmann ◽  
A. Osborne ◽  
A. Chiriano ◽  
...  

We studied the interchangeability of two blood gas syringes (Johns, Hardie Health Care Products Pty Ltd and Marksman, Martell Medical Products Inc) for the collection of blood for the analysis of PCO2, PO2, pH, sodium, potassium and glucose in 71 intensive care unit patients. The interchangeability of these two syringes with a specially designed syringe (Radiometer, Radiometer A/S) for the collection of blood for the analysis of ionized calcium was also studied. Analysis of pH, sodium, potassium and glucose showed no clinically significant differences between samples collected with Johns and Marksman syringes. However, differences in PCO2 and PO2 in samples collected with these syringes may be clinically significant if the PO2 is less than 100 mmHg. There were no clinically significant differences in ionized calcium levels in blood samples collected with Johns, Marksman and Radiometer syringes. We conclude that Johns and Marksman syringes are interchangeable for the collection of blood for the analysis of PCO2, PO2, pH, sodium, potassium and glucose and they are also interchangeable with Radiometer syringes for the collection of blood for ionized calcium analysis.


Blood ◽  
2006 ◽  
Vol 108 (11) ◽  
pp. 4007-4007
Author(s):  
Lisa J. Wakeman ◽  
Roger C. Munro ◽  
Saad Al-Ismail ◽  
Ann Benton ◽  
Andrew Beddall ◽  
...  

Abstract Introduction: The recommended order of draw for multiple tube collections (NCCLS [CLSI] H3-A5) clearly indicate that citrate tubes for coagulation tests should be taken before any other tubes (except blood cultures) and that a discard tube should be used if specialised coagulation tests are to be performed. This is to avoid the possibility of tissue activation and the theoretical risk of additive carryover on the parameters being measured. Because of the paucity of published evidence, this study was performed to determine the effect of the order of draw and whether the use of a discard tube is really necessary. Methods: Three consecutive early morning venous samples were collected into siliconised glass B–D Vacutainers containing tri-sodium citrate (Ref: 367691) from 116 healthy laboratory personnel (F= 74; M = 42) aged 20–63 yrs. Age groups were equally represented. Samples were processed on a Sysmex CA1500 analyser within 1 hour of collection. Appropriate CLSI guidelines were followed throughout. All parameters were measured using Dade-Behring reagents: Activated partial thromboplastin time (APTT) (Actin FSL), prothrombin time (PT) and derived fibrinogen (DF) (Innovin), thrombin clotting time (TCT) (Thromboclotin) and Clauss fibrinogen (CF) (Bovine thrombin and Owren’s veronal buffer). For each parameter, the data from each of the three samples were analysed for significant differences by one way analysis of variance (ANOVA). Results: Data obtained on measurements of basic coagulation parameters are shown in the table below. SDs are shown in parenthesis. (ns = not significant). Coagulation Parameter Results and Statistical Analysis Parameter First Sample Second Sample Third Sample ANOVA (p) ns=not significant APTT (secs) 28.3 (1.73) 28.3 (1.73) 27.9 (1.64) 0.230 (ns) PT (secs) 10.9 (0.47) 10.9 (0.47) 10.8 (0.45) 0.368 (ns) TCT (secs) 15.8 (1.03) 15.8 (1.02) 15.7 (1.02) 0.740 (ns) DF (g/L −1) 2.44 (0.54) 2.47 (0.55) 2.48 (0.55) 0.866 (ns) CF (g/L −1) 3.03(0.67) 3.04 (0.67) 3.10 (0.67) 0.825 (ns) No statistically significant differences were found between the first, second or third samples for any of the measured parameters. Conclusions: The CLSI recommends an order of draw for evacuated blood collection tubes in order to reduce the possibility of tissue activation in coagulation samples and the theoretical risk of additive carryover on the parameters being measured. Until now, this was based largely on theoretical probability. This comprehensive study demonstrates that the use of a discard tube is probably unnecessary since there is no statistical difference in any of the parameters measured between the first, second or third samples. Although this potentially obviates the expensive use of a discard tube in normal subjects, further work is required to determine whether it is necessary when measuring abnormally prolonged parameters in various pathological states.


Blood ◽  
2010 ◽  
Vol 116 (21) ◽  
pp. 3348-3348
Author(s):  
Cassandra Josephson ◽  
John Roback ◽  
Robert Myers ◽  
Lisa Hallock ◽  
Irene DeMezzo ◽  
...  

Abstract Abstract 3348 Background: Technologies have recently been developed for rapid determination of extended human erythrocyte antigen (xHEA) phenotypes. For example, a semi-automated method using allele-specific oligonucleotides targeted against 32 clinically significant minor RBC antigens has been used to determine donor xHEA phenotypes from whole blood samples. This approach is currently used by blood collection centers and medical centers with blood collection facilities (both sites have access to linked donor whole blood samples). Broader access to xHEA information closer to the point-of-care (e.g. Transfusion Services at a Medical Center without a blood collection facility) may provide an opportunity to enhance patient care by more quickly and broadly providing units with xHEA phenotypes (Klapper et al., 2010.) However, transfusion services would need to use integrally attached segments for testing, and with leukoreduced (LR) RBC units these segments have very low numbers of white blood cells (WBC) (and therefore DNA), potentially limiting analysis. This study was performed to determine whether a HEA-elongation mediated multiplex assay in solution (HEA-eMAP-S) (Xin et al., 2010) could accurately genotype segments from LR-RBC units for 32 clinically significant minor RBC antigens. Methods: Segments from pre-storage LR-RBC units (American Red Cross), < 14 days old, were obtained from a large tertiary care Children's Hospital in the Southeastern US and residual WBC were quantified by flow cytometry. DNA was extracted using an extraction method developed at BioArray SolutionS (BAS) using commercial reagents (Qiagen, Inc., Valencia, CA), and then amplified with the Universal Beadchip™ package (HEA LR-eMAP-S Beadchip™ Kits) which contains allele specific oligonucleotides directed to 32 clinically significant blood group antigens (c, C, e, E, V, VS, K, k, Kpa, Kpa, Jsa, Jsb, Jka, Jkb, Fya, Fyb, M, N, S, s, Lua, Lub, Dia, Dib, Coa, Cob, Doa, Dob, Joa, Hy, Yta, Ytb mutation for hemoglobin S). DNA analysis results were correlated with RBC storage solution, WBC filter type, and serologic minor RBC antigen phenotypes of the units. Results: 102 LR-RBC units from whole blood donations were studied, 74 /102 (73 %) stored in AS-1 and 28 /103 (27 %) in CPDA-1 solution. All AS-1 units were pre-storage LR with Fenwal Sepacell Flex Excel Filters and all CPDA-1 units were pre-storage LR with Whole Blood Fenwal Filters (Fenwal Inc. Lake Zurich, IL). All units demonstrated < 5 × 106 WBC/unit with 47 % having < 4 × 104 WBC/unit, which is at or below the limit of flow cytometric detection. Complete genotyping data was obtained from all samples. Ten samples showing initial indeterminate results on Diego and one for N antigens produced complete results after repeat testing. Fifty-four percent of units were serologically phenotyped for 1–8 antigens by the blood collection center; there was 100% correlation between predicted phenotype from DNA analysis and serology for these units. Conclusions: The HEA LR-eMAP-S DNA analysis can be applied to optimally pre-storage LR-RBC units yielding > 99 % accuracy for all minor red blood cell antigens tested. The ability to perform this type of testing in a hospital transfusion service opens up new possibilities for transfusion services to select from their existing inventory and more efficiently allocate units to recipients with specific phenotypic requirements for RBC units. Disclosures: Josephson: Immucor: Speakers Bureau. DeMezzo:Immucor: Employment. Tanzi:Immucor: Employment. Enriquez:Immucor: Employment. Lin:Immucor: Employment. Hashmi:Immucor: Employment.


Author(s):  
Merve Ergin ◽  
Serpil Erdogan ◽  
Onur Akturk ◽  
Ozcan Erel

AbstractBackground:This research investigated the effects of the transport of blood samples between centers/laboratories by car on coagulation tests.Methods:Five tubes of blood samples were taken from 20 healthy volunteers. The samples consisted of a baseline (control) group, centrifuged and noncentrifuged transported samples; centrifuged and noncentrifuged untransported samples. The groups of centrifuged and noncentrifuged samples were transported by car for 2 h. The centrifuged and noncentrifuged untransported samples were incubated in the laboratory until the transported samples arrived. Prothrombin time (PT) and activated partial thromboplastin time (APTT) tests were conducted for all samples.Results:Significant differences between the baseline group and the centrifuged and noncentrifuged transported samples and the noncentrifuged untransported samples were found for APTT levels (p<0.05, for all). In addition, significant mean percentage differences in PT values were found between the baseline group and the noncentrifuged transported samples (p<0.001) and the noncentrifuged untransported samples (p=0.005). The mean level of PT in the noncentrifuged transported samples was outside the upper limit of the clinical decision level.Conclusions:Noncentrifuged transported samples showed clinically significant differences in PT test results that may have stemmed from mechanical agitation during transportation. Therefore, we recommend not transporting noncentrifuged specimens for PT testing by car.


1984 ◽  
Vol 30 (4) ◽  
pp. 553-556 ◽  
Author(s):  
J Toffaletti ◽  
N Blosser ◽  
K Kirvan

Abstract We studied the stability of ionized calcium and pH in samples stored at either room temperature or 4 degrees C, in centrifuged and uncentrifuged blood-collection tubes and in centrifuged tubes containing a silicone-separator gel (SST tubes). At room temperature, in uncentrifuged blood from healthy individuals, mean ionized calcium usually increased no more than 10 mumol/L per hour; at 4 degrees C it did not change detectably for 70 h. This stability was fortuitous, however: the concentrations of both hydrogen and lactate ions in these samples increased, apparently with offsetting effects on the concentration of ionized calcium. Blood stored for 70 h at 4 degrees C in centrifuged SST tubes, although showing a slightly greater change in ionized calcium, had less change of pH and no change in the ionized calcium corrected to pH 7.4. In 11 heparinized whole-blood samples from eight patients in intensive care, the mean change per hour in ionized calcium and pH after storage at room temperature was +10 mumol/L and -0.04 units, respectively.


2018 ◽  
Vol 64 (5) ◽  
pp. 782-790 ◽  
Author(s):  
Mads Nybo ◽  
Merete E Lund ◽  
Kjell Titlestad ◽  
Christian U Maegaard

Abstract BACKGROUND Pneumatic transportation systems (PTSs) are increasingly used for transportation of blood samples to the core laboratory. Many studies have investigated the impact of these systems on different types of analyses, but to elucidate whether PTSs in general are safe for transportation of blood samples, existing literature on the subject was systematically assessed. METHODS A systematic literature review was conducted following the preferred reporting items for systematic reviews and metaanalyses (PRISMA) Statement guidelines to gather studies investigating the impact of PTS on analyses in blood samples. Studies were extracted from PubMed and Embase. The search period ended November 2016. RESULTS A total of 39 studies were retrieved. Of these, only 12 studies were conducted on inpatients, mainly intensive care unit patients. Blood gases, hematology, and clinical chemistry were well investigated, whereas coagulation, rotational thromboelastometry, and platelet function in acutely ill patients were addressed by only 1 study each. Only a few parameters were affected in a clinically significant way (clotting time parameter in extrinsic system thromboelastometry, pO2 in blood gas, multiplate analysis, and the hemolysis index). CONCLUSIONS Owing to their high degree of heterogeneity, the retrieved studies were unable to supply evidence for the safety of using PTSs for blood sample transportation. In consequence, laboratories need to measure and document the actual acceleration forces in their existing PTS, instituting quality target thresholds for these measurements such as acceleration vector sums. Computer modeling might be applied to the evaluation of future PTS installations. With the increasing use of PTS, a harmonized, international recommendation on this topic is warranted.


2019 ◽  
Vol 45 (1) ◽  
pp. 26-36 ◽  
Author(s):  
Arzu Kösem ◽  
Canan Topçuoğlu ◽  
Sevilay Sezer ◽  
Şimal Köksal Cevher ◽  
Ezgi Coşkun Yenigün ◽  
...  

Abstract Objective Blood collection tubes (BCTs) related interferences in test results can adversely influence on patient outcomes. We compared test results of samples in BD (Becton-Dickinson, Franklin Lakes, NJ, USA) Vacutainer Serum Separator Tubes (SST), BD Vacutainer® Barricor™ Plasma BCTs (Barricor™) and BD Vacutainer® Rapid Serum Tube (RST). Materials and methods Thirty-two samples were obtained from patients after the hemodialysis were included in this study. Eight routine clinical chemistry parameters (AST, creatinin, urea, PTH, glucose, LDH, K, calcium) were measured on Roche Cobas Analyzer (Roche Diagnostics, North America). The results of samples obtained from RST and Barricor™ were compared with SST as reference tubes. Results Results of Glucose, K, Urea, PTH from the SST and Barricor™ were statistically significantly different (p = 0.017, p < 0.001, p = 0.011, p < 0.001, respectively). In addition, results of PTH, LDH from SST and RST were significantly different (p < 0.001, p = 0.019). However, statistical significance of test results was not clinically significant for the biochemical parameters. Conclusion Working with Barricor™ may provide not just a fast, clean, high-quality plasma samples, safety results, but also time and cost-effectivity. Therefore, these types of tubes, which are less costly than other BCTs, may be preferred to obtain plasma.


Sign in / Sign up

Export Citation Format

Share Document