A Note about the Individualized TMS Focality
AbstractA particular yet computationally successful solution of an inverse transcranial magnetic stimulation (TMS) problem is reported. The goal has been focusing the normal unsigned electric field at the inner cortical surface and its vicinity (the D wave activation site) given a unique high-resolution gyral pattern of a subject and a precise coil model.For 16 subjects and 32 arbitrary target points, the solution decreases the mean deviation of the maximum-field domain from the target by a factor of 2 on average. The reduction in the maximum-field area is expected to quadruple. The average final deviation is 6 mm.Rotation about the coil axis is the most significantly altered parameter, and the coil moves 10 mm on average during optimization. The maximum electric field magnitude decreases by 16% on average. Stability of the solution is enforced. The relative average de-focalization is below 1.2 when position/orientation accuracies are within 3 mm/6 degrees, respectively. The solution for the maximum normal field may also maximize the total field and its gradient for neighboring cortical layers III-V (I wave activation).