Comparison between traditional and alternative biocompatible welding techniques used in orthodontic devices

2020 ◽  
pp. 146531252097240
Author(s):  
Fernanda de Souza Henkin ◽  
Luciane Macedo de Menezes ◽  
Berenice Anina Dedavid ◽  
Cátia Abdo Quintão

Objective: To compare the mechanical strength of joints made by conventional soldering with those made by alternative, more biocompatible, methods (spot, tungsten inert gas [TIG] and laser welding), and to compare the microstructural morphology of wires welded with these techniques. Design: In vitro, laboratory study. Methods: Forty stainless-steel wire segments with 0.8-mm diameter were joined by silver soldering, spot, laser and TIG welding. Ten specimens were produced for each one. Tensile strength test was performed 24 h after welding on the Emic DL2000™ universal testing machine, using a load cell of 1000 N with a crosshead speed of 10 mm/min. Results: The highest tensile strength mean values were obtained with silver soldering (532 N), next were laser (420 N), spot (301 N) and TIG (296 N) welding. Statistically significant differences were observed between the groups; the Dunn post-hoc test revealed differences between laser and spot welding ( p=0.046), laser and TIG ( p = 0.016), spot and silver ( p <0.001), and silver and TIG ( p <0.001). Conclusion: Laser welding strength is high, and comparable to silver welding. Spot and TIG techniques present comparable and significantly lower strengths. The four methods presented resistance values compatible with orthodontic use. The microstructural morphology is different for each technique. The association between the mechanical performance and the microstructure evaluation shows that laser presented the highest quality joint.

Author(s):  
Yan Zhang ◽  
DeShui Yu ◽  
JianPing Zhou ◽  
DaQian Sun ◽  
HongMei Li

Abstract To avoid the formation of Ti-Ni intermetallics in a joint, three laser welding processes for Ti alloy–NiTi alloy joints were introduced. Sample A was formed while a laser acted at the Ti alloy–NiTi alloy interface, and the joint fractured along the weld centre line immediately after welding without filler metal. Sample B was formed while the laser acted on a Cu interlayer. The average tensile strength of sample B was 216 MPa. Sample C was formed while the laser acted 1.2 mm on the Ti alloy side. The one-pass welding process involved the creation of a joint with one fusion weld and one diffusion weld separated by the remaining unmelted Ti alloy. The mechanical performance of sample C was determined by the diffusion weld formed at the Ti alloy–NiTi alloy interface with a tensile strength of 256 MPa.


2008 ◽  
Vol 78 (5) ◽  
pp. 917-921 ◽  
Author(s):  
Paola Gandini ◽  
Linda Orsi ◽  
Chiara Bertoncini ◽  
Sarah Massironi ◽  
Lorenzo Franchi

Abstract Objective: To test the hypothesis that there is no difference between the frictional forces produced by a passive self-ligating bracket (SLB) in vitro and a conventional bracket (CB) used with two types of elastomeric ligatures. Materials and Method: The brackets, wires and ligation methods used in vitro were a passive SLB and a CB used with two types of elastomeric ligatures (conventional elastomeric ligature [CEL] and unconventional elastomeric ligatures [UEL]). The bracket ligation systems were tested with two types of wires (0.014″ super elastic nickel titanium wire and 0.019″ × 0.025″ stainless steel wire). Resistance to sliding of the bracket/wire/ligature systems was measured with an experimental model mounted on the crosshead of an Instron testing machine with a 10 N load cell. Each sample was tested 10 consecutive times under a dry state. Results: Frictional forces close to 0 g were recorded in all tests with SLB and in all tests with UEL on CB with both wire types. Resistance to sliding increased significantly (87–177 g) (P &lt; .05) when CEL on CB was used with both wires. Conclusion: UELs may represent a valid alternative to passive SLBs for low-friction biomechanics.


2010 ◽  
Vol 31 (11) ◽  
pp. 1124-1129 ◽  
Author(s):  
Saima Aslam ◽  
Rabih O. Darouiche

Objective.To examine the mechanical integrity of hemodialysis catheters after exposure to the combination of N-acetylcysteine, ti-gecycline, and heparin, compared with that of catheters exposed to heparin alone.Methods.We used 3 types of hemodialysis catheters: polyurethane, silicone, and carbothane catheters. Catheter segments were incubated in vitro for various time intervals of up to 2 weeks either in a novel catheter lock solution (novel CLS), which consisted of N-acetylcysteine, tigecycline, and heparin, or in heparin alone (as a control). At the time of testing, each segment was rinsed and cut longitudinally into 2 sections. All catheter sections were scanned using an optical dissecting microscope to check for surface abnormalities and to measure wall thickness. We also carried out tensile strength testing of another set of catheters using a universal testing machine. Tested parameters included stress at yield, strain at yield, stress at break, strain at break, modulus of elasticity, and force at break.Results.The surfaces of catheters in both groups appeared similar by microscopy. The mean thickness of the catheter wall was not significantly different for the catheters exposed to the novel CLS and the catheters exposed to heparin (P >.05). Results for most of the tensile strength parameters tested were similar in the 2 groups of catheters at the end of 2 weeks of incubation. In particular, the force-at-break value of all tested catheters remained much greater than that recommended by industry standards.Conclusions.The use of the novel catheter lock solution did not impair the mechanical integrity or increase the propensity for fracture of hemodialysis catheters.


2019 ◽  
Vol 7 (13) ◽  
pp. 2162-2166 ◽  
Author(s):  
Rasha M. Abdelraouf ◽  
Manar Mohammed ◽  
Fatma Abdelgawad

AIM: This study aimed to assess the shear bond strength of a self-adhering flowable resin composite versus a total-etch one to different surfaces of permanent-molars. MATERIAL AND METHODS: Thirty-six sound human permanent molars were used. The teeth were embedded in acrylic blocks, such that their buccal surfaces were shown. The teeth were divided into three groups: Group I: Uncut-Enamel, Group II: Cut-enamel-surfaces with minimal-grinding and Group III: dentin-surfaces. Half of the teeth in each group were used for bonding to a self-adhering flowable resin-composite (Dyad-flow, Kerr, USA). While the other half of each group was bonded to a total-etch flowable resin-composite (Filtek™Z350-XT,3M-ESPE, USA) which necessitate etching and bonding. Teflon-mold was used for constructing resin composite cylinders (3 × 3 mm) over the buccal surfaces. The Dyad-flow was applied in the central hole of the mould placed upon tooth-surface, and then light-cured for 20 seconds. The Filtek-Z350-XT was applied similarly after etching and bonding steps. The teeth were stored in 37°C distilled water for 24 hours. The strength was measured using a universal testing machine and statistically analysed. Modes of failure were studied using digital-microscope. RESULTS: Mean values of shear bond strength for the Dyad and Filtek-Z350-XT in the uncut-enamel were 3.5 and 24.6MPa respectively, while that for cut-enamel were 4.5 and 12.7MPa respectively (Both highly statistically significant P ≤ 0.01) and in dentin were 4.3 and 6.7MPa respectively (Statistically significant P ≤ 0.05). The failure mode for Dyad was mainly adhesive (un-cut or cut-enamel 83.3% adhesive and 16.7% mixed, while in dentin 100% adhesive). While the modes of failure for Filtek-Z350-XT in enamel, either cut or un-cut, were 50% cohesive and 50% mixed, whereas in dentin 100% adhesive. CONCLUSION: Bonding of self-etch ″Dyad-flow″ flowable resin-composite was lower than the total-etch one in enamel and dentin. Thus further material improvement may be required.


2021 ◽  
pp. neurintsurg-2020-017242
Author(s):  
David Fernandez-Sanchez ◽  
Daniel Garcia-Sabido ◽  
Tudor G Jovin ◽  
Helena Villanova ◽  
Tommy Andersson ◽  
...  

BackgroundANA Advanced Neurovascular Access provides a novel funnel component designed to reduce clot fragmentation and facilitate retrieval in combination with stent-retrievers (SRs) in stroke patients by restricting flow and limiting clot shaving. In previous publications ANA presented excellent in vitro/in vivo efficacy data, especially with fibrin-rich hard clots. We aimed to determine the main physical property responsible for these results, namely suction force versus aspiration flow.MethodsWe evaluated in a bench model the suction force and flow generated by ANA and compared them to other neurovascular catheters combined with a SR (Solitaire). Aspiration flow was evaluated with a flow rate sensor while applying vacuum pressure with a pump. Suction force was determined using a tensile strength testing machine and a purposely designed tool that completely seals the device tip simulating complete occlusion by a hard clot. Suction force was defined as the force needed to separate the device from the clot under aspiration. All experiments were repeated five times, and mean values used for comparisons.ResultsAspiration flow increased with the inner diameter of the device: ANA 1.85±0.04 mL/s, ACE68 3.74±0.05 mL/s, and 8F-Flowgate2 5.96±0.30 mL/s (P<0.001). After introducing the SR, the flow was reduced by an average of 0.57±0.12 mL/s. Due to its larger distal surface, ANA suction force (1.69±0.40 N) was significantly higher than ACE68 (0.26±0.04 N) and 8F-Flowgate2 (0.42±0.06 N) (P<0.001). After introducing the SR, suction force variation was not relevant except for ANA that increased to 2.64±0.41 N.ConclusionDespite lower in vitro aspiration flow, the ANA design showed a substantially higher suction force than other thrombectomy devices.


2019 ◽  
Vol 53 (2) ◽  
pp. 117-125
Author(s):  
Jayanti Choudhary ◽  
B Shashikumar ◽  
Anand K Patil

Aims: This study aimed to evaluate and compare the effect of tea tree oil (TTO) mouthwash and chlorhexidine (CHX) mouthwash on frictional resistance. Settings and Design: In vitro. Materials and Methods: In total, 60 extracted premolars were mounted on a custom-made acrylic fixture. These 60 premolars were randomly divided into 3 groups of 20 each, on which 0.022″ × 0.028″ slot MBT stainless steel brackets were bonded and 0.019″ × 0.025″ rectangular stainless steel wire was ligated with an elastomeric module. The 3 groups included a control group where the samples were immersed in artificial saliva and 2 experimental groups immersed in 0.2% CHX and TTO mouthwash, respectively, for 1.5 hours. Postimmersion static frictional resistance was evaluated on a universal testing machine at crosshead speed of 0.5 mm/min. Statistical Analysis Used: Tukey’s post hoc procedure. Results: This study showed a statistically significant difference in the frictional resistance between saliva and CHX groups and CHX and TTO groups ( P < .05). No statistically significant difference was observed between saliva and TTO groups ( P > .05). The frictional resistance was more in the CHX mouthwash group than in the TTO mouthwash group. Conclusions: Frictional resistance was lesser in the TTO mouthwash than in the CHX mouthwash. Based on this result, TTO mouthwash can be used instead of CHX mouthwash as an oral hygiene aid in patients with orthodontic treatments.


2013 ◽  
Vol 24 (6) ◽  
pp. 626-629 ◽  
Author(s):  
Maria Isabel A. Faria ◽  
Erica Alves Gomes ◽  
Danielle Cristine Messias ◽  
Joao Manoel Silva Filho ◽  
Celso Bernardo Souza Filho ◽  
...  

The aim of this in vitro study was to evaluate the tensile strength of glass fiber posts submitted to different surface treatments. Forty-eight maxillary canines had their crowns sectioned and root canals endodontically treated. The roots were embedded in acrylic resin and distributed into 3 groups according to the surface treatment: Group I: the posts were treated with silane agent for 30 s and adhesive; Group II: the posts were cleaned with alcohol before treatment with silane agent and adhesive; Group III: the posts were submitted to conditioning with 37% phosphoric acid for 30 s before treatment with silane agent and adhesive. Each group was divided into 2 subgroups for adhesive polymerization or not before insertion into the canal: A - adhesive was not light cured and B - adhesive was light cured. All posts were cemented with Panavia F and the samples were subjected to tensile strength test in a universal testing machine at crosshead speed of 1 mm/min. Data were submitted to one-way ANOVA and Tukey's test at 5% significance level. There was statistically significant difference (p<0.01) only between group GIII-B and groups GI-A and GI-B. No significant difference was found among the other groups (p>0.05). It was concluded that the products used for cleaning the posts influenced the retention regardless of adhesive light curing.


2012 ◽  
Vol 23 (1) ◽  
pp. 49-53 ◽  
Author(s):  
Neilor Mateus Antunes Braga ◽  
Juliana Monteiro da Silva ◽  
Jacy Ribeiro de Carvalho-Júnior ◽  
Raquel Conceição Ferreira ◽  
Paulo César Saquy ◽  
...  

This in vitro study compared different ultrasonic vibration modes for intraradicular cast post removal. The crowns of 24 maxillary canines were removed, the roots were embedded in acrylic resin blocks, and the canals were treated endodontically. The post holes were prepared and root canal impressions were taken with self-cured resin acrylic. After casting, the posts were cemented with zinc phosphate cement. The samples were randomly distributed into 3 groups (n=8): G1: no ultrasonic vibration (control); G2: tip of the ultrasonic device positioned perpendicularly to core surface and close to the incisal edge; and G3: tip of the ultrasonic device positioned perpendicularly to core surface at cervical region, close to the line of cementation. An Enac OE-5 ultrasound unit with an ST-09 tip was used. All samples were submitted to the tensile test using an universal testing machine at a crosshead speed of 1 mm/min. Data were subjected to one-way ANOVA and Tukey's post-hoc tests (α=0.05). Mean values of the load to dislodge the posts (MPa) were: G1 = 4.6 (± 1.4) A; G2 = 2.8 (± 0.9) B, and G3= 0.9 (± 0.3) C. Therefore, the ultrasonic vibration applied with the tip of device close to the core's cervical area showed higher ability to reduce the retention of cast post to root canal.


2015 ◽  
Vol 2015 ◽  
pp. 1-8 ◽  
Author(s):  
Dino Re ◽  
Francesco De Angelis ◽  
Gabriele Augusti ◽  
Davide Augusti ◽  
Sergio Caputi ◽  
...  

Purpose.Although new elastomeric impression materials have been introduced into the market, there are still insufficient data about their mechanical features. The tensile properties of 17 hydrophilic impression materials with different consistencies were compared.Materials and Methods.12 vinylpolysiloxane, 2 polyether, and 3 hybrid vinylpolyether silicone-based impression materials were tested. For each material, 10 dumbbell-shaped specimens were fabricated (n=10), according to the ISO 37:2005 specifications, and loaded in tension until failure. Mean values for tensile strength, yield strength, strain at break, and strain at yield point were calculated. Data were statistically analyzed using one-way ANOVA and Tukey’s tests (α=0.05).Results.Vinylpolysiloxanes consistently showed higher tensile strength values than polyethers. Heavy-body materials showed higher tensile strength than the light bodies from the same manufacturer. Among the light bodies, the highest yield strength was achieved by the hybrid vinylpolyether silicone (2.70 MPa). Polyethers showed the lowest tensile (1.44 MPa) and yield (0.94 MPa) strengths, regardless of the viscosity.Conclusion.The choice of an impression material should be based on the specific physical behavior of the elastomer. The light-body vinylpolyether silicone showed high tensile strength, yield strength, and adequate strain at yield/brake; those features might help to reduce tearing phenomena in the thin interproximal and crevicular areas.


2007 ◽  
Vol 8 (3) ◽  
pp. 21-28 ◽  
Author(s):  
Marjaneh Ghavamnasiri ◽  
Sara Abedini LNU ◽  
Alireza Mahdizadeh Tazangi

Abstract Aim The aim of this in vitro study was to evaluate the effect of 20% carbamide peroxide (CP) on the flexural strength (FS) of bovine enamel and dentin complex. Methods and Materials One hundred sound bovine teeth were selected for use. Blocks (2×3×8 mm) from the middle portion of the facial surfaces of each crown were sectioned from the teeth. The specimens were randomly divided into five groups (n=20) based on the time period of vital bleaching. Group 1 comprised the control group kept in artificial saliva. The experimental groups subjected to immersion in 20% CP, Opalescence, for four, six and eight weeks, respectively (eight hours daily). Mechanical testing was performed 24 hours after the last treatment using an Instron Universal Testing Machine with a crosshead speed of 0.5 mm/min. The results were analyzed by one-way analysis of variance (ANOVA) and Tukey's test (α=0.05). Results Mean values for FS were 141.27, 103.06, 120.71, 130.23, and 117.90 (Mpa), respectively. Statistically significant differences in FS were noted among the groups. A statistically significant difference was observed between Group 2 (two week bleach) and the control. Conclusion Application of 20% CP bleach for two weeks provided significant decreases in FS of bovine enamel and dentin complex. Similar decreases were not observed among the other groups. Citation Ghavamnasiri M, Abedini S, Tazangi AM. Effect of Different Time Periods of Vital Bleaching on Flexural Strength of Bovine Enamel and Dentin Complex. J Contemp Dent Pract 2007 March;(8)3:021-028.